3.2 Connecting Input/Output Cables
3.2.2 Connecting to CH2 Terminal
(9766 CLAMP ON SENSOR)
• In case of external noise, wind the cable around
the ferrite clamp supplied as an accessory as
shown below.
• Note that measurement may be adversely
affected by external noise or the electromagnetic
environment when using a cable longer than 3
meters.
9766 CLAMP ON SENSOR
CH2
Connect the extension cable from the 9766
CLAMP ON SENSOR to the CH2 terminal.
3.2.1, "Connecting to CH1 Terminal (voltage/
contact pulse signals input)" (page 22)
• Use only the 9766 CLAMP ON SENSOR for cur-
rent pulse input.
• The clamp sensor connection is irrespective of
polarity.
• Attach the clamp around only one conductor.
Single-phase (2-wire) or three-phase (3-wire)
cables clamped together will not produce any
reading.