5.2 Function Specifications
32
Measurement
accuracy
±100 ppm rdg. ±1 dgt.
Recording interval accuracy: ±2 ms
Display range: Total of each channel at every
interval (16,000,000 max.)
Guaranteed
accuracy period
1 year
Sampling
1 time / sec.
Input terminal
3 Input terminal block × 2
5.2 Function Specifications
Actual time
management
Time management using PC application
Recording
start/end
Recording is started and ended by a com-
mand. (Immediate start, timed start, or timed
end)
Recording
Method
Interval recording (Total number of pulse are
recorded at a set recording interval.)
Recorded data
One data set contains time, temperature,
and humidity (At every interval).
Recording mode Instantaneous value recording mode
(Total during the interval)
Recording end
condition
Memory full stop or indefinite
Set the mode before the start of recording.
Quantity of
recorded data
512 KB Flash memory
Instantaneous value recording mode (Total
during the interval) 30,000 data × 2 CH
Data acquisition
method
All logging data, data before, at, and after the
specified time, or the current instantaneous
value (monitored value)
Recording
interval
1/2/5/10/15/20/30 sec.
1/2/5/10/15/20/30/60 min.
Содержание 2304-20
Страница 2: ......
Страница 14: ...Notes on Use 10 ...
Страница 21: ...17 1 4 Dimension Diagrams 1 4 Dimension Diagrams 96 1 22 5 1 85 1 93 8 1 103 9 1 100 5 1 9 6 0 5 Unit mm ...
Страница 22: ...1 4 Dimension Diagrams 18 ...
Страница 24: ...2 1 Setting the Module ID 20 ...
Страница 40: ...6 2 Servicing 36 ...