17

© 2022 S. Himmelstein and Company—all rights reserved.     www.himmelstein.com

Examples:

Remember *ALL* messages to and from the MCRT48200V series end with a <CR> or a <LF>.

•  Retrieve data: 

Send “DC” to the MCRT48200V. The “DC” is the data current command. The return message should look 
something like “1234.56”.

•  Retrieve the filter: 

Send “FL” to the MCRT48200V. The return message should be something like “07” which implies (refer-
ring to the appropriate list under the “FL” message) the filter setting is 5 Hz.

•  Set the filter to 100 Hz: 

Refer to the list under the “FL” (filter) command to find that a 100 Hz filter corresponds to the value 02. 
Therefore, send “FL02” to the MCRT48200V. The MCRT48200V should respond with “OK” if the operation 
was successful.

•  Retrieve the version number of the MCRT48200V: 

Send “VR” to the MCRT48200V. The return message should be something like “Model MCRT48200V v1.2”.

Informational Only Messages

•  T

hese messages can only retrieve information from the MCRT48200V—they can not change any data on the 

MCRT48200V.

•  The time returned is the number of 10kHz clock ticks since the MCRT48200V was powered on.

MESSAGE

REPLY

MEANING

DC

float

Current engineering data for the torque .

XC

hex

Current hexadecimal data for the torque .

YC

hex, hex

Time and Current hexadecimal data for torque .

64

base64

Current base-64 data for the given channel .

65

hex, base64

Hexadecimal time and Current base-64 data for the given channel .

MX

int, int

The max and min of the torque . Multiply by the scaling constants (SC command) to get lbf-in .

SN

string

Serial number of the MCRT48200V .

SP

string

Scratch pad of the MCRT48200V .

MD

string

Model number of the MCRT48200V .

PP

Bool

Is flash password protected?

See the “PC” command to change the protection password, and the “PS” command to set/unset the 

protection .

VR

string

Version number of firmware .

AD

int, int, int

A/D readings

 

o

1st number: temperature sensor

 

o

2nd number: F-coil voltage

 

o

3rd number: Input voltage

SC

hexfloat, hexfloat

Positive/Negative Scaling Constants, lbf-in / counts

Содержание MCRT 48200V

Страница 1: ...NON CONTACT DIGITALTORQUEMETER INSTALLATION OPERATION ANDTROUBLESHOOTING GUIDE MCRT 48200V NON CONTACT DIGITALTORQUEMETER INSTALLATION OPERATION ANDTROUBLESHOOTING GUIDE COMPANY 2490 Pembroke Avenue H...

Страница 2: ...te Unless specified differently when ordered default values are as follows Sensor Full Scale 10 Volts Torque Unit of Measure lbf in Filter Cutoff Frequency 10 Hz Password Protection Off These selectio...

Страница 3: ...D 2 3 Starting High Inertias with Electric Motors 10 D 3 Allowable Bearing Loads 11 D 4 Allowable Extraneous Loads 11 D 4 1 Allowable Bending Loads 11 D 4 2 AllowableThrust Loads 11 D 5 Operating Spee...

Страница 4: ...tude variations That data is then transmitted to the stator as infrared IR energy where it is detected and output in industry standard analog and digital formats The IR transmission is inherently tole...

Страница 5: ...force Next allow the assembly to cool to room temperature Then repeat the process for the second coupling If desired use forced air to accelerate cooling Air cooling avoids contaminating the torque m...

Страница 6: ...ailable at www belden com Himmelstein recommends the connec tion in Figure 20 For best results use a differential input amplifier in these electrically noisy environments B 2 2 Power Input 2 Pin Conne...

Страница 7: ...1 Zero Velocity Code Z Pickup Code Z Zero Velocity Speed Pickups have three pin connectors They require external power 6 to 15VDC Connections are as follows Pin A Power In 6 to 15VDC Pin B Output Sig...

Страница 8: ...pe A Pin A Signal Pin B Signal Type Z Pin A POWER IN Pin B Output Signal Pin C Common Notes 1 A dot appears next to Pin 1 on all connectors 2 All cables are shielded 3 Shields Float at Torquemeter end...

Страница 9: ...c data with unit of measure Displays Torque Peak and Valley numeric data with unit of measure Displays Torque Spread numeric data with unit of measure Classifies status of user set High and Low Torque...

Страница 10: ...taneous torque and the torquemeters overload rating as a safety margin for unexpected loads Do not knowingly operate in the overload region If you expect torques in the overload region then change to...

Страница 11: ...inches of bending or 1 of the applied pounds of thrust D 4 1 Allowable Bending Loads When it is applied without thrust a standard MCRT Torquemeter mounted as a floating shaft can handle a shaft bendi...

Страница 12: ...ntial faults The best procedure is to isolate the problem part then correct or replace it Otherwise return the defective part to the factory E 2 Preliminary Inspection E 2 1 Torquemeter Inspect the se...

Страница 13: ...1 Hertz If the readings become steady then you may wish to identify the physical cause of the shaft torque variation or remove it with mechanical filtering techniques Oscillographic signal analysis i...

Страница 14: ...nominal Has reverse polarity and 25 Volt overvoltage protection Available Options Enhanced Performance Code C Code N is standard performance 60 PPR Speed Pickup Options Code A is standard type Code Z...

Страница 15: ...or aType A pickup and 3 pins for aType Z pickup To order spare connectors use the following part numbers P N When the bearing load ratings don t meet the above criteria use pillow blocks and a jack sh...

Страница 16: ...it mantissa with an implied MSB of 1 0x80000000 Sign 0 positive 1 negative 0x7F800000 Exponent 0x007FFFFF Mantissa o if sign 0 then value 2 Exponent 24 Mantissa 0x800000 o else value 2 Exponent 24 Man...

Страница 17: ...se messages can only retrieve information from the MCRT48200V they can not change any data on the MCRT48200V The time returned is the number of 10kHz clock ticks since the MCRT48200V was powered on ME...

Страница 18: ...ut E DAC value to give 10V analog output Miscellaneous Messages MESSAGE REPLY MEANING OK Write data in RAM to non volatile flash PS index string OK Change Flash password protection o string must match...

Страница 19: ...the user Only data from the DC command is subject to this scaling FL int FiLter 0 10 4 bits o 2 200 Hz o 3 100 Hz o 4 50 Hz o 5 20 Hz o 6 10 Hz o 7 5 Hz o 8 2 Hz o 9 1 Hz o 10 0 5 Hz o 11 0 2 Hz o 12...

Страница 20: ...COMPANY 2490 Pembroke Avenue Hoffman Estates IL 60169 847 843 3300 www himmelstein com CDTM...

Отзывы: