Page 34
187692(10) - EN - P280 Mattress Instructions for Use
For all other disinfectants :
Proceed with the final steps before reusing the cleaned and disinfected mattress
(See “Performing the Final Steps” page 35).
Cleaning and Disinfecting at Regular Intervals or in the Event of
High Risk of Contamination
Follow the same instructions as above (Voir “Cleaning and Disinfecting after the
Departure of a Patient or when Changing the Sheets” page 33), but with the
products listed below.
Using Recommended Products
After the departure of patients with infectious diseases:
•
Dismozon
®
Pur
(Bode)
Concentration: 40 g/l (= 4%), contact time: 1 hour
After the departure of patients with Clostridium Difficile Infection:
•
Dismozon
®
Pur
Concentration: 15 g/l (= 1.5%), contact time: 2 hours
•
Anioxy-Spray WS (Anios)
Ready to use, contact time: 30 min
Dry Steam Cleaning the Mattress
Dry steam or superheated steam contains no more than 6% of water in
suspension and avoids the effects of condensation.
To avoid damage due to high pressure or an abnormal surface temperature, take
the following precautions:
•
Use low pressure steam on the electric parts.
•
Do not use accessories such as
high pressure hoses (A). It is
preferable to use soft non-metallic
brushes (B) and wipers (C) in such a
way as to reduce the pressure to an
acceptable level.
•
Prevent water and steam from
getting into connectors that are not
in use.
•
Do not brush and apply reduced pressure to labels and markings.
•
Dry carefully and look for signs of water ingress.
•
Test the device before reusing it.
Chlorine-based solutions can be used:
Concentration must be less than or equal to 1,000 ppm.
Ethanol-based solutions (alcohol) can be used:
Concentration must not exceed ¼ of ethanol for ¾ of water.