ALARMS, ALERTS, AND TROUBLESHOOTING
7
ALARMS, ALERTS, AND TROUBLESHOOTING
7
NOTES:
• Options for resolving the alert are based on the configuration of the ventilation system.
• When attempting to resolve an alert, ensure that the patient receives adequate ventilation therapy; place the
patient on an alternate means of ventilation if necessary.
COMPRESSOR ALERTS
The compressor has alerts that are independent of the ventilator. Compressor alerts must be resolved in order for
the compressor alert notifications to be silenced; there is no Silence Alarm button for alerts originating from the
compressor.
NOTIFICATION
CAUSE
CHECKS AND POSSIBLE RESOLUTION IN EACH CONFIGURATION
STAND-ALONE CONFIGURATION
EXTENDED RANGE CONFIGURATION
STATIONARY CONFIGURATION
Low Battery Alert
(Intermittent buzzer)
Compressor battery capacity drops
to 20% or less.
N/A
Connect the compressor to an AC power source.
If the battery does not recharge, place the
patient on an alternate means of ventilation
(if necessary) and contact your service
representative.
Connect the compressor to an AC power source.
If the battery does not recharge, place the patient on an
alternate means of ventilation (if necessary) and contact your
service representative.
Constant Audible Alert
(Compressor stops
operating)
Multiple causes: high motor
temperature, high electronics
temperature, electronic circuit
error, or motor stall.
N/A
Discontinue use of the compressor. Place the
patient on an alternate means of ventilation
(if necessary), or connect the ventilator to an
alternate pressure source (oxygen cylinder or
wall source) in Stand-Alone Configuration.
If running on battery, check the compressor’s
battery charge status. If the status is less than
two indicator lights, connect the compressor to
an AC power source.
Power off the compressor and power it on again.
If the alert persists, power off the compressor
and contact your service representative.
Discontinue use of the compressor. Place the patient on an
alternate means of ventilation (if necessary), or undock the
ventilator from the compressor and connect to an alternate
pressure source (oxygen cylinder or wall source) in Stand-
Alone Configuration.
If running on battery, check the compressor’s battery charge
status. If the status is less than two indicator lights, connect
the compressor to an AC power source.
Power off the compressor and power it on again.
If the alert persists, power off the compressor and contact
your service representative.