41
42
STAND-ALONE CONFIGURATION
4
STAND-ALONE CONFIGURATION
4
POLE MOUNT
You can attach the ventilator to a pole using an optional pole mount. For information about ordering accessories
and replacement parts, see "Accessories and Replacement Parts" on page 147.
CAUTION:
• Make sure the pole mount is securely attached to the pole, and the ventilator and clip are securely
fastened to the pole mount and the ventilator. If the clip is not securely attached to the pole mount and the
ventilator, the ventilator may fall and be damaged.
• Always secure the ventilator to prevent it from falling or becoming damaged.
• Use only the approved belt clip and pole mount with the ventilator.
Position the pole mount around the pole in the correct orientation.
Turn the knob on the pole mount until the pole mount is securely
attached to the pole.
Slide the belt clip for the ventilator into the hole on the top of the
pole mount and push down until it is secure.
Attach the ventilator to the belt clip on the pole mount by lining
up the belt clip with the belt clip sockets on the ventilator. Push
the ventilator in towards the vent clip and pole mount until the
ventilator audibly clicks into place.
VENTILATOR SILENCE ALARM BUTTON
Silencing and clearing alarms is a multi-step process that depends on alarm priority and how many alarms are
active. For more information see "Chapter 7: Alarms, Alerts, and Troubleshooting".
Silence Alarm button on ventilator.
Press the Silence Alarm button to temporarily silence the alarm for 60
seconds. Pressing the Silence Alarm button silences only one alarm
at a time—in audible or vibrating alarm mode. If more than one alarm
occurs, press the Silence Alarm button once for each alarm. If the
alarm is a medium- or high-priority alarm and is not silenced after 60
seconds, the alarm will continue with an additional buzzer.
Resolve the condition that triggered the alarm. For help resolving
possible causes of an alarm and options to resolve it. If a Silence Alarm button is pressed but the condition
that triggered the alarm is not resolved, the alarm will sound again after 60 seconds.
After resolving a High Temperature, High Circuit Pressure, or High PEEP
Pressure high-priority alarm, touch
OK
in the message that indicates the
alarm has been resolved.
1
500 ml
High Circuit
Peak Pressure
has been resolved.
10:11a, Fri
Jun 12, 2015
3
Values displayed on the screen are for
illustrative purposes only.
POWERING OFF SEQUENCE FOR THE VENTILATOR
To power off the ventilator, press the Ventilator Power button
for three seconds until a confirmation screen appears.
To continue to power off the ventilator, choose
OK
.
NOTE:
If a selection is not made within 20 seconds or if the
BACK
button is
selected, the previous screen will be displayed and the ventilation status will
not be affected.
1