7.0 Preventive Maintenance, Multirall™ 200 overhead lift
Part
Part No.
Replacement
interval:
Replacement:
Hand Control
3136001
2 years
Can be used more than 2 years if no malfunction is detected.
Battery
31290007
3 years
Can be used more than 3 years if no malfunction is detected.
Lift Strap
31390014
5 years
Must be replaced after 5 years of use, or if wear and tear are detected.
Expected Life Time
Normal use is defined as 8-12 normal lifts/day
Year
Multirall™ 200 overhead lift
10
Expected Life Time
The expected life time of a Multirall™ overhead lift is 10 years of normal use provided that required
Periodic Inspection and Preventive Maintenance is performed by personnel authorized by Hill-Rom
and using original Liko™ spare parts.
However certain parts have a shorter life cycle and will need to be replaced in order for the overhead lift
to meet its expected life time. They are listed below: