REPLACING THE ELECTROMAGNET

STEP

5

STEP

5

STEP

4

STEP

4

STEP

6

STEP

6

STEP

7

STEP

7

This  completes  the  electromagnet  replacement 

procedure.

Insert the wire into the silicon tube, and tie up in a 

bundle. 

Strip the coating off each wire. 

(5-7mm from the end) 

Connect the wires to the terminals. 

Solder the wires to the terminals. 

The wire requires a soldered connection.

Mount new electromagnet in the unit. 

Fix frame stay temporarily. 

(Do  not  tighten  screw  of  frame  stay  and  keep 

them rather loose.) 

Tighten up hexagonal nuts with nut driver. 

Fasten screws of frame stay.

MOUNTING THE ELECTROMAGNET

Install the SP switch to the frame with the screws.

 

●Be careful of the direction of switch lever. (Refer to the 

pictures)

8

HP

Replacing the Electromagnet

Содержание hp-150

Страница 1: ...er Cover Semi Cover Packing Filter Fitting Boss Upper Housing Sound Absorber Lap Casing A Valve Chamber Packing Casing B Valve Diaphragm Ring Diaphragm Diaphragm Base Actuating Rod SP Switch Safety pi...

Страница 2: ...ul permanent magnets Therefore be sure to remove your watch and any other precision machines before operation as they may be affected by the strong magnetic force Do not put the actuating rod close to...

Страница 3: ...t driver to loosen or tighten the U lock nut STEP 7 STEP 7 Fit a new diaphragm and diaphragm ring in the diaphragm base Be sure not to leave a gap between them Remove one of the diaphragm mounting blo...

Страница 4: ...n fix them with 5 screws STEP 11 STEP 11 Fit the actuating rod by aligning it with the groove and tighten U lock nut and flat washer by the nut driver Use new U lock nut and washer otherwise U lock nu...

Страница 5: ...ning boss on the diaphragm base into the concave part of the frame stay Secure the diaphragm mounting block on the other side and tighten washers and U lock nuts with the nut driver Make sure that gap...

Страница 6: ...b of the actuating rod is even Be careful not to touch the terminal when the power is on testing the operating conditions as this will result in an electric shock Unplug the pump immediately after the...

Страница 7: ...the hexagonal nuts by the nut driver 8mm wrench Loosen the screw holding the frame stay and remove it Pull out the electromagnet from the pump body CAUTION Be sure to unplug the pump unit When perform...

Страница 8: ...he wires to the terminals Solder the wires to the terminals The wire requires a soldered connection Mount new electromagnet in the unit Fix frame stay temporarily Do not tighten screw of frame stay an...

Отзывы: