IRP40.50
28
2.5
Aligning Stylus to Spindle Centre
1. Loosen clamping screws C (Fig. 8) (4x) then slightly retighten.
2. Adjust the probe with the adjustment screws A (Fig. 8) (4x) to within less than 20µm.
3. Tighten the clamping screws C (Fig. 8) (4x) a little more.
4. Adjust the probe with the adjustment screws A (Fig. 8) (4x) to within less than 5µm.
5. Tighten the clamping screws C (Fig. 8) (4x).
6. Tighten the adjustment screws A (Fig. 8) (4x) against each other.
7. Check the alignment.
8. Calibrate the probe.
Fig. 8
Aligning stylus to spindle centre
C = Clamping screw AF1.5 mm (4x)
M= 1.5 Nm (1.11 lbf.ft)
(5362)
Hexagon key
AF1.3 mm (5370)
AF1.5 mm (5369)
A = Adjustment screw
DIN913 M2.5x5 AF1.3 mm (4x)
M= 1.0 Nm (0.74 lbf.ft)
(5363)
Dial gauge
(3079)