LS-R-4.8
31
2.7 Maintenance and Cleaning
2.7.1 Cleaning the optical windows of the laser scanner
NOTICE
Risk of measuring errors!
Damage to the optical windows affects the measuring accuracy of the laser scanner.
•
Contact the service department at m&h Inprocess Messtechnik GmbH.
NOTICE
Risk of measuring errors!
Grease and dirt deposits on the optical windows reduces the measuring accuracy of the sensor.
•
Do not touch the optical windows of the sensor with bare hands or contaminate it otherwise.
•
Remove contamination from the optical windows before measuring.
•
Do not use aggressive cleaning agents, as they may damage the optical windows.
1. Wipe the optical windows with a clean, soft, lint-free micro-fibre cloth or with the supplied optics cleaning cloth
(4297).
2. If the contamination cannot be removed with the cloth, gently wipe the windows with a camera lens cleaner or
isopropyl alcohol.
3. If the contamination cannot be removed even with a camera lens cleaner or isopropyl alcohol, contact the
service department at m&h Inprocess Messtechnik GmbH.
Cleaning the optical windows of the laser scanner
Optical windows of the
laser scanner