Hettich AG, Seestrasse 204a, 8806 Baech, Switzerland
Operator Manual_DancerPlus_EN_rev, 1.1
Page 18
Fig. 3: Rack R5123
Fig. 4: Rack R0010-FAST
1
4
1
5
3
Fig. 12: Rack R0121
2
Legend:
1
Hole for the temperature sensor
2
Holes (4 pieces) for mounting to the shaking plate
3
Special holes for the FAST rack pins on the shaking plate
4
Mounted handles, optional CD.Z909
5
FAST securing base for the FAST pins on the shaking plate with additional distancing from the shaking
plate, e.g. for the gripper arm of a robot (CD.Z908 for 30mm height, or CD.Z912 for 60mm height)
The R0121 rack for the DWP (3 pieces) or microtiter plates are not suitable for the FAST option
for the shaking plate.
When the rack is removed, the temperature sensor must first be removed from the hole and laid
aside in order to prevent the PTFE sleeve of the temperature sensor or its cable from being bent
or scraped.
If the temperature sensor is bent, the sensor itself may become broken and the device will no
longer function as a result.
If the PTFE sleeve is scraped, this may cause a leakage which will prevent the vacuum pump
from being able to reach the final value. In both cases, the guarantee on a new temperature
sensor, its installation and potential transport costs will become void.