Argyll II

™ User/Service Manual 

 

MAINTENANCE 

 
Transportation 

 

Care should be taken when loading, unloading and during transportation & refrain 
from  placing heavy objects on top of the mattresses. 

 

Do not expos to direct sunshine for any period of time when loading and unloading & 
ensure a cover is used to protect the mattress when raining or snowing. 

 

Do not open with a knife or sharp objects as this may damage the mattress cover. 

 

Always use the provided re-usable pump protector for all transportation methods. This plastic 
sleeve will need to be cleaned prior to re-use for infection control purposes. 

 

Calibration 

  During service intervals or between patients, it is recommended to check pressure 

tolerances (see the table on page 9) using a Herida Testing System. Calibration training 
must be conducted & applied by a Herida trained technician. 

 

Other Considerations 

 

Original internal fuse is F0.5A 250V. When replacing the fuse, ensure a similar type is 
used and that  the fuse is disposed of in the appropriate manner. 

 

During normal working, when the emergency deflation is needed, switch off the pump 
and pull out the CPR toggle quickly. 

 

The mattress should be away from sharp objects to avoid any scratches or punctures. 

 

Put bed cover sheet or blanket over the mattress before use. No plastic or rubber 
sheets should be used. 

 

Ensure that all the tubes remain flat & any twists / kinks are avoided. 

 

If any air leakage is found, check all the tube connectors and air tubes carefully to see if 
any  are loose or damaged. 

 

If any major problems or issues are found or the product is not working, stop 
using  immediately and ask the engineer / manufacturer / provider for assistance. 

 

Any discarded mattress or covers should be treated in accordance with local rules 
and  guidance in accordance with contaminated products. 

 

The pump is factory set to 80Kg and is a guide only. This can be adjusted depending 
on the needs of the patient.  

 

 

Parts and Accessories

 

 

Mattress (1PC) / Pump

1PC) / Air tube

1PC) / Mattress cover

1PC

 

 
 
 

Содержание ARG-121-2

Страница 1: ...ervice Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattres...

Страница 2: ...2 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual...

Страница 3: ...OPERATION 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair...

Страница 4: ...ght Do not use phenolic based cleaning solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags s...

Страница 5: ...under patient weight as is often the case Application This product is to be used for those patients who have existing tissue damage or pressure ulcers It can also be used as a high risk preventative i...

Страница 6: ...gly to appropriately accommodate see below Unless the clinical condition dictates it is not advised to adjust the bed profile beyond 45 The Argyll II will continue to operate beyond this but may cause...

Страница 7: ...on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting the...

Страница 8: ...Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed th...

Страница 9: ...tance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 50...

Страница 10: ...n trained by Herida Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This...

Страница 11: ...is should be done via a pressurised system soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear...

Страница 12: ...fuse is F0 5A 250V When replacing the fuse ensure a similar type is used and that the fuse is disposed of in the appropriate manner During normal working when the emergency deflation is needed switch...

Страница 13: ...uding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is sou...

Страница 14: ...14 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual P09 3 Way Pump...

Страница 15: ...gainst electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electric...

Страница 16: ...tors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the us...

Страница 17: ...1HE Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of in an envir...

Отзывы: