37
21. PRECAUTION
1. Place the analyzer a flat and rigid surface.
2. Electric power should be stable.
3. Analyzer and Cartridge are not to be exposed to direct sunlight.
4. Follow the indicated Analyzer operating temperature (17 - 32°C) and Cartridge
storage temperature (2 - 32°C).
5. Do not move or interfere with the analyzer during testing.
6. The bead inside the Reagent Solution of the Reagent Pack, might sink and
coagulate.
Gently
mix the Reagent Pack before testing, and visually check that the
beads are not coagulated.
7. Allow the test cartridge to reach room temperature (ambient temperature) 30
minutes before use.
8. When collecting blood, touch the end of the capillary tip (sampling area) to the
blood sample, do not put the tip deep into the blood sample to avoid excess blood.
9. Ensure that the cartridge compartment is empty before inserting the cartridge.
10. Insert the Cartridge gently in the Analyzer.
11. Firmly insert the reagent pack into cartridge with a mild pressure
until a snap is heard or felt. Too much force may affect the mixture of the two
reagents resulting in an incorrect result.
GB
Содержание HbA1c 501
Страница 1: ...Operating Manual HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 2: ......
Страница 43: ...41 MEMO GB ...
Страница 45: ...Bedienungsanleitung HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 46: ......
Страница 87: ...41 MEMO DE ...
Страница 89: ...Gebruiksaanwijzing HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 90: ......
Страница 131: ...41 MEMO NL ...
Страница 133: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 134: ......
Страница 175: ...41 MEMO SE ...
Страница 177: ...Brugsanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 178: ......
Страница 219: ...41 NOTATER DK ...
Страница 221: ...Bruksanvisning HbA1c 501 Analyzer GB DE NL SE DK NO ...
Страница 222: ......
Страница 263: ...41 NOTATER NO ...