25
13. SYSTEM CHECK
IMPORTANT
•
Do not open the analyzer lid during testing.
•
Do not move the analyzer during testing.
•
Do not expose the analyzer to vibration during testing.
6) The test result is displayed after 5 minutes.
NOTE:
The test result should be within
the target range specified on the package
insert of the control solution.
7) Take out the cartridge after the test is completed.
Possible causes of control solution results out of specified target:
•
Deterioration of the cartridge due to high humidity, heat, or
overexposure to light.
•
Deterioration of the control solution.
Action:
•
Repeat the QC test using a fresh cartridge from a new box or new lot. If
the result still is out of range, proceed to next possible cause.
•
Repeat the QC test with a fresh bottle of control solution.
•
Contact your local distributor for support.