43
Technical specifications (EMC-RF)
Use only cables with the following specification:
USB shielded maximum 2 m
Serial shielded maximum 1.5 m
Guidance and manufacturer´s declaration – electromagnetic emissions
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or the user of the HemoCue systems
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
Group 1
The HemoCue systems uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
not likely to cause any interference in nearby electronic
equipment.
RF emissions
Class B
The HemoCue systems are suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
NOTE 1
It is the manufacturer’s responsibility to provide equipment electromagnetic compatibility information to the customer or user.
NOTE 2
It is the user’s responsibility to ensure that a compatible electromagnetic environment for the equipment can be maintained in order that
the device will perform as intended.
Содержание Hb 301
Страница 1: ...HemoCue Hb 301 Operating manual Manuel d utilisation ...
Страница 2: ......
Страница 4: ...4 ...
Страница 6: ...QRG CD 1 2 3 4 5 6 Composants 6 Components ...
Страница 8: ...1 2 3 4 5 Mise en service 8 Start up ...
Страница 10: ...1 2 3 Configuration Signal sonore 10 Set up Audio signal ...
Страница 13: ...13 ...
Страница 14: ...1 2 3 4 Mesure Sang capillaire 14 Measuring Capillary blood ...
Страница 16: ...5 6 7 8 16 ...
Страница 18: ...9 10 11 18 ...
Страница 20: ...12a 12b 13 14 15 20 ...
Страница 24: ...1 2 3 Configuration Fonction PC Impression 24 Set up PC Printer function ...
Страница 26: ...1 2 3 4 4a Maintenance 26 Maintenance ...