41
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
See NOTE 2 and NOTE 3
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the HemoCue
systems, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d=1.2
√
P
d=1.2
√
P 80 MHz to 800 MHz
d=2.3
√
P 800 MHz to 2.5 GHz
Where (P) is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and (d) is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (a),
should be less than the compliance level in each
frequency range (b).
Interference may occur in the vicinity of equipment
marked with the following symbol.
Содержание Hb 301
Страница 1: ...HemoCue Hb 301 Operating manual Manuel d utilisation ...
Страница 2: ......
Страница 4: ...4 ...
Страница 6: ...QRG CD 1 2 3 4 5 6 Composants 6 Components ...
Страница 8: ...1 2 3 4 5 Mise en service 8 Start up ...
Страница 10: ...1 2 3 Configuration Signal sonore 10 Set up Audio signal ...
Страница 13: ...13 ...
Страница 14: ...1 2 3 4 Mesure Sang capillaire 14 Measuring Capillary blood ...
Страница 16: ...5 6 7 8 16 ...
Страница 18: ...9 10 11 18 ...
Страница 20: ...12a 12b 13 14 15 20 ...
Страница 24: ...1 2 3 Configuration Fonction PC Impression 24 Set up PC Printer function ...
Страница 26: ...1 2 3 4 4a Maintenance 26 Maintenance ...