32
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
See NOTE 2 and
NOTE 3.
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer
to any part of the HemoCue systems,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=1.2
√
P
d=1.2
√
P 80 MHz to 800 MHz
d=2.3
√
P 800 MHz to 2.5 GHz
Where (P) is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and (d) is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey (a), should be less than
the compliance level in each frequency
range (b).
Interference may occur in the vicinity of
equipment marked with the following
symbol.
Содержание Glucose 201
Страница 1: ...HemoCue Glucose 201 with Plasma Conversion Operating Manual ...
Страница 4: ...4 Components 4 2 1 QRG 3 5 6 ...
Страница 6: ...6 Start up 1a 2 3 4 5 1b ...
Страница 8: ...8 1 2a 2b 3 Audio signal ...
Страница 10: ...10 Measuring Capillary blood 1 2 3 4 ...
Страница 12: ...12 8 7 6 5 ...
Страница 14: ...14 9 10 11 ...
Страница 16: ...16 14 15 12 14 13 ...
Страница 18: ...18 3 1 2a 2b 4 Measuring Venous and arterial blood ...
Страница 20: ...20 2 3 4 Maintenance Daily maintenance 1 ...
Страница 22: ...22 5b 5c Maintenance Optical parts 5a 6 ...
Страница 36: ......