25
Symptom
Explanation
Action
Seen during start up and is a fatal error.
1. The analyzer cannot detect a valid EEPROM
memory configuration.
911
1a. Turn off the analyzer and turn it on again after
30 seconds and retry testing.
1b. The analyzer needs service. Call your distributor.
Seen during start up and is a fatal error. Self test of RAM
memory failed.
1. The electronics are out of order.
913
1. The analyzer needs service. Call your distributor.
Seen at start up and is a fatal error. Calibration checksum is
not valid.
1. The analyzer needs calibration.
925
1. The analyzer needs service. Call your distributor.
1. Communication error during internal hardware test.
929
1. The analyzer needs service. Call your distributor.
1. The electronic “SELFTEST” failed.
930
1a. Turn off the analyzer and allow it to reach room
temperature.
1b. If the problem continues, the analyzer needs service.
Call your distributor.
1. Measured value exceeds 444 mg/dL (24.6 mmol/L).
HHH
1. Check the expiration date of the microcuvettes.
2. Remeasure sample with fresh microcuvette.
1. The analyzer is not receiving power.
2. If on battery power, the batteries need to be replaced.
3. The display is out of order.
No characters on the display.
1a. Check that the power adapter is connected to the
power source.
1b. Check that the power adapter is securely connected
to the analyzer.
1c. Check that the adapter wire is not damaged.
2. Turn off the analyzer and replace the batteries,
4 type AA.
3. The analyzer needs service. Call your distributor.
1. The display is out of order.
2. The microprocessor is out of order.
The display gives erroneous
characters.
1. The analyzer needs service. Call your distributor.
2. The analyzer needs service. Call your distributor.
Содержание Glucose 201
Страница 1: ...HemoCue Glucose 201 with Plasma Conversion Operating Manual ...
Страница 4: ...4 Components 4 2 1 QRG 3 5 6 ...
Страница 6: ...6 Start up 1a 2 3 4 5 1b ...
Страница 8: ...8 1 2a 2b 3 Audio signal ...
Страница 10: ...10 Measuring Capillary blood 1 2 3 4 ...
Страница 12: ...12 8 7 6 5 ...
Страница 14: ...14 9 10 11 ...
Страница 16: ...16 14 15 12 14 13 ...
Страница 18: ...18 3 1 2a 2b 4 Measuring Venous and arterial blood ...
Страница 20: ...20 2 3 4 Maintenance Daily maintenance 1 ...
Страница 22: ...22 5b 5c Maintenance Optical parts 5a 6 ...
Страница 36: ......