Helmer Scientific Automatic Cell Washing System - UltraCW II
Operation Manual
360386/D
16
6 Product Specifications
6 .1
Operating Standards
These units are designed to operate under the following environmental conditions:
♦ Indoor use only.
♦ Altitude (maximum): 2000 m.
♦ Ambient temperature range: 5 °C to 40 °C.
♦ Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C.
♦ Transient over-voltages typically present on the MAINS supply.
♦ Do not exceed maximum loaded weight (including tubes and contents). Maximum weight is printed on rotor.
Table 3. Unit Specifications
UltraCW
®
II
Physical
Height
11” (280mm)
Width
13” (330mm)
Depth
18.9” (480mm)
Weight
53.8 lb (24.4 kg)
Rotor Radius
4.1” (105mm)
Maximum RPM / RCF
3500 RPM / 1438 RCF
Speed Tolerance
± 20 RPM
Kinetic Energy
250 Nm
Maximum Density
1.2 kg/dm
3
Noise Level
‹
49db
Electrical
Electrical Power (AC)
100V - 240V, 50/60 Hz
Electrical Power (DC)
6A @ 24 VDC
Fuses
10A / 250V
Maximum Current Consumption
2.2 A
Power Source
150W