On-Board FastRegen Control Module Programming and Operation Instructions
P/N 8040406
C-5
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
W
DELAY RESTART
This countdown timer delays starting the pump for cooldown mode
after it has been warmed up, roughed out, and passed the ROR test.
This is a user programmed value. If the pump has no valves, the delay
timer begins when the pump reaches 310K.
To get the timer value in minutes, use the command k.
NOTE: Display the time in hours to match the keypad display.
X, Y
POWER FAILURE
There was a power failure during the regeneration cycle and the pump
will make a decision to cooldown or regenerate. Refer to
Appendix D for more information.
Z
DELAY START
NOTE: Applies to Full and Fast Regeneration cycles.
The start of regeneration will be delayed for some time. To get the
timer value in minutes, use command k.
NOTE: Display the time in hours to match the keypad display.
O, [
ZEROING TC GAUGE
The TC gauge is being zeroed at the end of regeneration.
f*
SHARE REGEN WAIT
Waiting to lock/acquire shared use of rough valve for multi-pump
Fast regenerations when using the Network Terminal and the rough
valve interlock.
e*
REPURGE
This is the special repurge cycle in Fast regeneration that initiates
after the pump has failed to rough to the rough test pressure within the
customer programmed time.
Table C-1: RS-232 REGEN Responses and Descriptions (Continued)