Appendix B - On-Board Module RS-232 Interface Protocol Format
B-14
P/N 8040406
C
HELIX TECHNOLOGY CORPORATION
-
TI
CRYOGENICS
v
None
Inquires pump for several regeneration flag conditions com-
bined as bits in one response. The bits are biased by "@" for
ASCII printable results. The bit weights are:
01H - Waiting for rough valve.
02H - Purge gas failure detected in regeneration.
04H - Heater failure detected in regeneration
08H - Fast regeneration has recovered from purge gas failure.
10H - Fast regeneration crossed over to a full regeneration.
20H - True when fast regeneration has started.
z
[0,1, ?]
Set/Clear/Query the user keypad access lockout state. When
True, the user is limited to the use of the MONITOR and
SERVICE functions on the keypad/display.
[B
1 - 2
Restore relay 1 or 2 to automatic settings from manual mode.
Table B-1: On-Board Module RS-232 Commands (Continued)
Command
Parameter
Description