- 48-
Revision 4.6 3/2004
Section 22.3
Labeling Function – Standard Data
Logging Mode
The label option allows the input of an alphanumeric
label of up to 16 characters in length. Multiple labels can
be used. For example, a label can be entered and data
taken to be stored under the label. After completion of
the measurement activities to be included under the fi rst
label another label may be entered and the next data set
will follow that label in the data storage and transmis-
sion sequence. Label information can be viewed on the
HHC system display and is transmitted when stored
values are reviewed via the computer upload program
detailed in Section 24.
To enter a label follow the steps below.
Step 1
With the HHC system on and reading pressure/
temperature, press the data log key on the HHC system
keypad.
DATA
LOG
In response the HHC system display will respond:
THE LAST DATA LOG MENU SCREEN USED
Step 2
Using the up/down and right/left arrows browse
through the data log menu and locate the word “label”.
The menu line that contains the label function appears as:
Label Erase
Step 3
Using the left/right arrow keys select the word
Label. When selected the word label will fl ash.
Label Erase
Step 4
With the word Label fl ashing press the enter key.
ENT
The HHC system will respond by displaying:
Enter Label
- - - - - - - - - - - - - - -
Step 5
Use the Up/down arrows to enter letters, num-
bers or a blank and the right/left arrows to move the
cursor to the position in which the character is to be
entered. When the label information is displayed press
the enter key.
ENT
The label is now stored and will appear in sequence with
the corresponding data logged values when the stored
information is viewed on the display or uploaded to a
computer.
Section 22.4
Automatic Data Logging –
Standard Data Logging Mode
The automatic data logging function allows for the
automatic, unattended logging of measurement data. The
data logging process commences immediately after the
completion of the set up of the automatic data function
to be performed. Automatic data logging can be accom-
plished by the following procedure.
Step 1
With the HHC system on and reading pressure
press the data log key on the keypad.
DATA
LOG
The HHC system will respond by displaying:
THE LAST DATA LOG MENU SCREEN USED.
Step 2
Using the up/down and right/left arrows browse
through the data log menu and locate the word “Auto”
The menu line with the word “Auto” appears as:
Auto Manual
To initiate automatic, time interval, data logging select
the word “Auto” by using the arrow keys. When selected
the word “Auto” will fl ash on the display. For manual,
operator controlled, data entry proceed to Section 22.5.
Auto Manual
Step 3
With the word “Auto” fl ashing press the enter key.
ENT
The HHC system will respond by displaying:
Enter Interval
sec. xxxxxx
Step 4
Using the number keys enter the desired time
interval in seconds. The interval may be from 0.1 sec-
onds to 65,000 seconds.
Enter Interval
sec. 12345
Step 5
After the desired time interval is shown on the
display press the enter key.
ENT
Step 6
The HHC system will respond by displaying the
following:
Readings to Log
None Left Rght Both
Step 7
Using the arrow key select the desired Quick