![Heinzmann Triton CPM 500 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/heinzmann/triton-cpm-500/triton-cpm-500_manual_2120033010.webp)
4 Operation
CPM 500 Electronic Peak Pressure Indicator
10
4.5 Charge battery
The CPM 500 has a re-chargeable battery. The battery
charging will start automatically after the CPM 500 unit
has been connected via USB cable from USB port to PC
(fig. 4,5,6,7)
Fig. 4: USB-port of CPM 500
Note: At first usage please connect CPM 500 via USB cable to PC.
Charge battery for 2h continuously without disconnecting from PC !
Fig. 5: CPM 500 connected via USB cable to PC
Fig. 6: Indication during charging process
Fig. 7: Indication charging process completed
Do not change battery during charging process while CPM 500 is connect via USB cable to PC.
When the indicator is connected by USB, the battery charging runs automatically. The indicator can detect the
charging status and continues charging until the battery is fully loaded. Then the charging process stops
automatically. It is not recommended to interrupt the charging process before it is automatically stopped. The
charging process is finished when the display message "CPM 500 charging..." has disappeared.
If the battery status is low and the charging process stops after 5-10 minutes, then the user can manually force
the unit to charge the battery for 90 minutes by pressing both arrows
+
at the CPM 500 unit
for 2 seconds (Forced Charging Mode).
Note:
It is not recommended to use the Forced Charging Mode when the battery charge condition is more than
40%, due to the risk of overcharging. Please see chapter 5.2 for a description how to check the charge state.
USB por