Overpressure safety relief device
........................................
.......................................................................23-30-01-04-98
11.4 Replacing the overpressure safety relief device at the condenser
Round bottom flask 500 ml
, coated .....................................
......................................................................... 514-83000-02
➨
Remove joint clip at the catchpot.
➨
Unscrew the four Torx screws at the counter holder of the
condenser and remove condenser. Thereby remove the
adapter from the inlet of the condenser.
➨
Pull off old overpressure safety relief device and install
new one. Check for correct position of the PTFE-foil under
the overpressure safety relief device.
➨
Reassembly in reverse order.
➨
Position adapter in the inlet of the condenser and screw
condenser with counter holder to the pump (Torx screws).