The controller itself is maintenance free.
Contamination of the pressure transducer or deposits will influence the accuracy of measurement.
Attention:
Never use a spiky or sharp-edged tool to clean the pressure transducer.
Clean a contaminated pressure transducer as follows:
➨
Disconnect vacuum connection.
➨
Fill the measurement chamber with a solvent (e. g. benzene) and allow sufficient cleaning time. Ob
-
serve all regulations concerning usage and disposal of solvents!
➨
Drain the solvent and dispose of in accordance with regulations, repeat cleaning if necessary.
➨
Rinse the gauge head chamber several times with alcohol in order to remove all solvent residues.
➨
Allow the pressure transducer to dry.
➨
Assemble vacuum connection.
➨
Readjust the pressure transducer if necessary.
10 Cleaning the pressure transducer
vacuum
connection
NOTICE