66
E
Clean and inspect pump heads
Only ROTAVAC vario control:
Use box-end spanner (SW 17) and loosen coupling ring of hose beside gas-ballast valve.
Remove hose fitting from hose with ¼ revolution with box-end spanner (SW 15). Do not remove
fittings from pump head.
¾
Leaks may be found on re-assembly.
ROTAVAC vario control, ROTAVAC vario tec und ROTAVAC valve tec:
Loosen 6 ea. straight hex-recessed screws on pump head and lift off cover ass'y. (cover, internal
part, and lid):
¾
Never attempt to loosen parts with sharp tools
(screwdriver), use rubber hammer or compressed air
instead.
To inspect valves, carefully remove cover from lid. Denote
location of valves and remove.
¾
Replace damaged valves.
Remove contamination with cleaning agent.
Inspect diaphragm for damage and replace on condition.
Explosion view: pump head
housing
bearing
diaphragm
back-up disc
diaphragm
diaphragm
cocking disc
with screw
cover
piston rod
valve
lid
Содержание ROTAVAC valve control