Commands:
query
range
reply
remark
action
LEDx1?
x1=0...7
LED=0000\r\n
LED is „OFF"
LED-status
LED=0001\r\n
LED is „ON"
DSP?
DSP=xxxx
xxxx=display
act. displays
command
range
action
TAx2!
x2=0...7
key
0=no key 1=Cal
2=start 3=sense of rotation
4=max 5=up
6=down 7=select
SPx3=xxxx!
x3=0 xxxx=0001..5999 (0.1 sec)
pause
x3=2 xxxx=0010..0599 (min)
x4=3 xxxx=0010..0750 (hour)
SDM=xxxx!
xxxx=000,0..9999 (0.1 sec)
volume, metered
SDZ=xxxx!
xxxx=0000..9999 (depends on transmission used)
rotational speed
SMM=xxxx!
xxxx=000,1..9999 (not during calibration)
ml/min
wait a minute for reply.
SSD=xxxx!
xxxx=0000..0014
hose #
general
reply
remark
action
OK\r\n
command successful
self explaining
ERROR\r\n
command failed
none
PO\r\n
metering complete
none
9.5. Recommended sizes of tubes
-
To achieve high metering accuracy, in particular in the metering by volume mode, we
recommend the smallest tube diameter available; higher metering accuracy of small dia.
tubes is a function of the higher number of revolutions of the pump drive.