Assembly
22
Configure Interfaces
➜
Configure interface to receive data in 7bit ASCII format.
➜
Each command string has to be finalized with CR (carriage return) and LF (line feed).
Interface parameters
9600 baud; parity: 1 even, 7 bits, 1 stop bit
Read command
Description
Output
IN_PV_1
Query actual temperature: probe temperature in °C
IN_PV_2
Query actual temperature: probe safety sensor in °C
IN_PV_3
Query actual temperature: hotplate in °C
IN_PV_4
Query actual temperature: hotplate safety sensor in °C
IN_PV_5
Query actual motor speed in rpm
IN_SP_1
Query set temperature probe / hotplate in °C
IN_SP_2
Query set temperature safety delta in °C
IN_SP_3
Query set motor speed in rpm
IN_MODE_2
Behavior after power cut
Write Command
Description
Output
OUT_SP_1 y
Enter set temperature probe / hotplate in °C
OUT_SP_2 y
Enter set temperature safety delta in °C
OUT_SP_3 y
Enter set motor speed in rpm
OUT_MODE_2 y
▪ 0 = all OFF after power return
▪ 1 = heating / stirring ON after power return
(set values remain valid)
▪ „o“
▪ „*“
Command
General commands: Description
Output
START_1(*) Start heater control, activate remote –
alternating display *Remote* (see footnote)
START_2(*) Start motor, activate remote –
alternating display *Remote* (see footnote)
STOP_1
Stop heater control
STOP_2
Motor speed=0 Motor-stop
RESET
Reset to initial state: heater off, motor off,
exit remote mode
STATUS
Request status
▪ 0 = ManualMode
▪ 1 = RemoteMode Start
(START_1 OR START_2)
▪ 2 = RemoteMode Stop
(STOP_1 AND STOP_2)
▪ <0 Error code in case of
fault
* Attention: control panel of device is disabled in remote mode