112
E
41. RS 232 interface commands
query
Command
action
remarks
Act commands
VAC?
Act Vacuum
VAP?
Act Vapor
BATH?
Act Bath
CHILL?
Act Chiller
SPEED?
Act Speed
TIME?
Act Time
elapsed time, continuous operation
COUNT?
Act Time
time, timer mode
Safety functions
SC?
Safety Cover
0 = action; 1 = no action
FS?
Flask Support
0 = action; 1 = no action
RS?
Rotation Stop
0 = action; 1 = no action
OP?
Over-
Temperature
Prot.
0 = action; 1 = no action
triggering
Command
action
remarks
Set commands
VAC=XXXX!
Set Vacuum
4-digit in mbar; range set 0000 to 1200
BATH=XXXX!
Set Bath
4-digit, in 1°C increments
range set 0000 to 1800 equals 0.0 to 180.0
SPEED=XXX!
Set Speed
3-digit, in rpm; range set 000 to 180
TIME=XXXX!
Set Timer
4-digit, in hh:mm
range set 0000 to 9959 equals 00:00 to 99:59
Actions
SVX!
Start Stop Vacuum
SV0! = OFF ; SV1! = ON
VVX!
Valve
vacuum valve; VV0! = CLOSE ; VV1! = OPEN
BVX!
Ventilation
aerating valve; BV0! = CLOSE ; BV1! = OPEN
SBX!
Start Stop Bath
SB0! = heater OFF ; SB1! = heater ON
SSX!
Start Stop Speed
SS0! = rotation OFF ; SS1! = rotation ON
LU!
lift UP
LD!
lift DOWN
Содержание 518-17100-00
Страница 1: ...LABOROTA 20 control automatic Betriebsanleitung Instruction Manual Mode d emploi...
Страница 2: ...2 BEDIENFRONT BASISGER T CONTROL PANEL BASIC SYSTEM PANNEAU FRONTAL DE COMMANDE APPAREIL DE BASE...
Страница 4: ......
Страница 26: ...26 D...
Страница 47: ...47 D...
Страница 107: ...107 E produkt sensor fitting cooling water separator aftercooler...
Страница 164: ......
Страница 167: ......
Страница 189: ......