4 Installation and connection
The controller is designed for installation directly at the workplace.
>
Observe all specifications for installation, connection and
operation according to technical data,
see chapter
10.1.1 Technical data
.
>
Also observe rating plate data.
Installation conditions
The controller has acclimatized.
Ambient conditions are observed and are within the limitation
of use.
Limitation of use
(US)
Ambient temperature
10
–
40 °C
50–104°F
Altitude, max.
3000 m
above sea level
9840 ft
above sea level
Relative humidity
30–85 %, non condensing
Degree of protection
(controller front)
IP 20 (IP 42)
Avoid condensation or contamination by dust, liquids or corrosive gases.
4.1 Installation
4.1.1
Table top version
The table top or bench-top type controller can be installed and
connected directly on top of the work bench or on laboratory table.
The table top version is supplied with a hose nozzle. The hose
nozzle should be positioned in a way that the connected vacuum
hose cannot kink.
Consider installation
conditions
Содержание 036308001
Страница 1: ...Betriebsanleitung Operating instructions Hei VAC Control Intelligent evaporation...
Страница 6: ...Erste Schritte First steps Neuger t Erste Schritte Neuger t First steps delivery status...
Страница 39: ...38 6 2 Sprache und Druckeinheit einstellen Sprache und Druckeinheit einstellen 5 Startbild...
Страница 77: ...Translation of the original instructions Page 82 151 Zertifikate Certifications Seite page 152 156...
Страница 81: ...81 First steps delivery status First steps on delivery status Select language and units...
Страница 114: ...114 6 2 Select language and pressure unit Select language and pressure unit 5 Initial screen...
Страница 124: ...124 6 7 Stop control Stop control 5 Controller and vacuum control stops 5 Display icons switched off...
Страница 142: ...142 Note the settings of stored programs before loading Default Load default settings IMPORTANT Load factory settings...
Страница 154: ......