![Heidolph 036308001 Скачать руководство пользователя страница 149](http://html1.mh-extra.com/html/heidolph/036308001/036308001_operating-instructions-manual_2118931149.webp)
149
Vacuum data
Hei-VAC Control, internal vacuum sensor
(US)
Measuring range, absolute
1080–0,1 mbar
810–0.1 Torr
Max. control range
1060–0,1 mbar
795–0.1 Torr
Resolution
0,1 mbar
0.1 Torr
Max. admissible media temperature (gas):
Temporarily
80 °C
176°F
Continuous operation
40 °C
104°F
Measurement uncertainty
< ±1 mbar
< ±0.75 Torr
Temperature coefficient
< ±0,07 mbar/K
< ±0.05 Torr/K
External vacuum sensor VSK 3000
Max. admissible pressure,
absolute
1,5 bar
1125 Torr
Venting
Max. admissible pressure,
absolute
1,2 bar
900 Torr
Gas connections
Hei-VAC Control table top
version
Fitting for PTFE tube 10/8 mm or
hose nozzle for flexible tube DN 6/10
Venting
Hose nozzle for flexible tube
d
i
= 4–5 mm
Display
Type
LC display (LCD)
Brightness control
yes
Pressure display
switchable: mbar, Torr, hPa
Weight and dimensions*
(US)
Weight (built-in)
440 g
0.97 lb
Weight with foot (table top)
570 g
1.3 lb
Dimensions
123 mm x 124 mm x 83 mm
5 in. x 5 in. x 3.5 in.
Dimensions with foot
144 mm x 124 mm x 115 mm
6 in. x 5 in. x 4.5 in.
* without wall power supply
Technical data
basic device
Содержание 036308001
Страница 1: ...Betriebsanleitung Operating instructions Hei VAC Control Intelligent evaporation...
Страница 6: ...Erste Schritte First steps Neuger t Erste Schritte Neuger t First steps delivery status...
Страница 39: ...38 6 2 Sprache und Druckeinheit einstellen Sprache und Druckeinheit einstellen 5 Startbild...
Страница 77: ...Translation of the original instructions Page 82 151 Zertifikate Certifications Seite page 152 156...
Страница 81: ...81 First steps delivery status First steps on delivery status Select language and units...
Страница 114: ...114 6 2 Select language and pressure unit Select language and pressure unit 5 Initial screen...
Страница 124: ...124 6 7 Stop control Stop control 5 Controller and vacuum control stops 5 Display icons switched off...
Страница 142: ...142 Note the settings of stored programs before loading Default Load default settings IMPORTANT Load factory settings...
Страница 154: ......