HRC Installation Instructions Vers. 001
7
WARNING Be sure to perform periodic data backup procedures. Check the success of the
backup to avoid data loss caused by backup errors.
WARNING To avoid the risk of electric shock, this equipment must only be connected to
a grounded power supply.
WARNING To avoid the risk of electric shock, do not touch conductive parts of
connectors and the patient simultaneously.
WARNING Do not operate the system directly after large temperature changes. Let the
device acclimate itself for a minimum of two hours to avoid device damage
or incorrect measurement results.
WARNING Clean and disinfect chin rest and headrest or other soiled parts after each
examination, Contaminated parts may lead to infections or disease
contraction.
CAUTION
In the rare event of defe
illumination slit, the live image will
appear as shown to the r
image shows a small streak in
horizontal direction on the monitor,
whereas the upper and l
the live image are dar
please turn the instrument off
immediately and contact Heidelber
Engineering or the local distributor.
Do not attempt to repai
instrument yourself.
CAUTION
Please evaluate all images concerning their image quality (e.g. brightness,
noises, artifacts) before making a diagnostic decision.
CAUTION
Please note that the distance or diameter values issued by the Imaging
software are only pixel values and not correlating to a linear dimension. The
real distances are depending on the camera setup, patient distance and
image quality.
CAUTION
If the camera or the objective were fallen off, please contact the Heidelberg
Engineering Service for a device check.
CAUTION
Please note that vibrations of the device location (e.g. caused by strong
machines) might lead to bad or distorted images.
CAUTION
Never leave the patient alone with the instrument during the examination!
CAUTION
The instrument must not be used under any circumstances if mechanical,
optical or electrical faults occur. Any change or addition to the system must
comply with the relevant legal guidelines. Repairs, particularly to the
electronic and optical components, must be carried out only by Heidelberg
Engineering authorized, trained personnel.
ctive
ight. The
ower parts of
k. In this case,
g
r the
Содержание HRC
Страница 4: ...HRC Installation Instructions Vers 001 4 ...
Страница 13: ...4 Hardware Installation HRC Installation Instructions Vers 001 13 ...
Страница 17: ...HRC Installation Instructions Vers 001 17 Remove the white plastic ring It is only necessary for transport ...
Страница 20: ...HRC Installation Instructions Vers 001 20 The head rest is screwed to the table from below ...