Weekly
Check that UV lamps light up during mobile dryer
operation.Also, check that all cables are undamaged. A
damaged cable can be a danger to life!
Quarterly
Change the air filter at least once per quarter. If the outer
surface of the filter becomes very dirty, it must be
changed sooner.
8. Filter replacement
1. Loosen the metal clips on the side plates of the
cassette.
2. Remove the used filter.
3. Insert the new filter in such a way that the IRT logo is
visible from the outside.
4. Push the long edges of the filter into the cassette
grooves.
5. Replace the metal clips.
Attention:
The filters are disposable and so should not
be cleaned as this will result in the loss of their filtering
capacity.
9. UV-lamp replacement
Please contact your distributor and order a UV-lamp
replacement kit including instructions.
After replacing the lamp a reset of the timer needs to be
done.
Reset Lamp timer:
Make sure that the main switch [5] is
turned off. Hold the Start/Stop button
[1] and top right button [2] pressed in
simultaneously while the main switch [5]
is turned on.
Keep the buttons pressed in until a
confirmation message (L1--rESEt.)
appears in the top display [4].
10. Trouble shooting
Problem
The UVA lamps starts up but shuts down automatically
after a while.
Possible solution
The lamp is worn out. Please contact your distributor and
order a new one.
Problem
The lamp doesn´t start.
Possible solutions:
—
Ensure that required cooling time for lamps is
fulfilled. Cooling time increases with dirty air filter.
— The lamp is worn out. Please contact your distributor
and order a new lamp.
— Filter glass is broken.
— No power in wall socket, check fuse.
[3]
[5]
[2]
[4]
[1]
11. Declaration of conformity
We, Hedson Technologies AB
Box 1530
SE-462 28 VÄNERSBORG Sweden
declare under our sole responsibility that the product
IRT-UVA 1/2 PrepCure 4 Digital
to which this declaration relates is in conformity to
the applicable provisions of US Federal requirements,
including the Code of Federal Regulations, CFR Title 21,
--FOOD AND DRUGS CHAPTER I--FOOD AND DRUG
ADMINISTRATION DEPARTMENT OF HEALTH AND
HUMAN SERVICES SUBCHAPTER J--RADIOLOGICAL
HEALTH.
UV-curing devices is not included in table 1 of 21
CFR 1002.1 and therefore not subject to a mandatory
radiation safety performance standard contained in title
21 CFR Part 1040 (optical radiation) but complies with
the general requirements in Title 21 CFR 1000 through
1005.
following the standards:
UL499 14th Ed. issued Nov. 7, 2014.
CSA C22.2 No.88 issued: 1958/09/01
Arlöv, Sweden, July 2016