Warnings and Cautions
Prior to use, the operator should read all warnings and cautions, instructions for
operation and maintenance r e q u i r e m e n t s contained in this operato
r’s m a n u a l .
WARNING:
DO NOT USE THE ODORATOR 2 OR BREATHE IN GAS FROM SOURCES WHICH MAY
CONTAIN TOXIC LEVELS OF CHEMICALS SUCH AS HYDROGEN SULFIDE (H
2
S) AS IT
COULD BE HARMFUL TO HUMAN HEALTH AND SAFETY TO DO SO. SOURCES WHICH
MAY CONTAIN TOXIC LEVELS OF CHEMICALS INCLUDE BUT ARE NOT LIMITED TO
PRODUCTION SYSTEMS AND LANDFILLS.
WARNING:
IT IS THE SOLE RESPONSIBILITY OF THE PRODUCT PURCHASER/ OWNER TO ENSURE
THAT: (1) ALL OPERATORS OF THE ODORATOR 2 READ, UNDERSTAND, AND COMPLY WITH
THE PRECAUTIONS, INSTRUCTIONS FOR OPERATION AND MAINTENANCE REQUIREMENTS
CONTAINED IN THIS MANUAL; (2) THE ODORATOR 2 IS ONLY USED IN A PROPER AND SAFE
MANNER; AND (3) WHILE USING THE ODORATOR 2, OPERATORS ARE NOT INHALING OR
SNIFFING DANGEROUS MATERIALS THAT COULD CAUSE INJURY OR DEATH. IF THE
EQUIPMENT IS USED IN A MANNER NOT SPECIF
IED BY THE USER’S MANUAL, THE PROTECTION
PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.
WARNING:
DURING OPERATION, KEEP THE ODORATOR 2 AWAY FROM OPEN FLAMES. INLET SUPPLY
PRESSURE SHOULD NOT EXCEED FOUR PSIG. INLET PRESSURES MORE THAN FOUR
PSIG C OULD RUPTURE THE LOW-PRESSURE REGULATOR DIAPHRAM, VENT THE GAS
OUT OF THE INSTRUMENT AND CAUSE INSTRUMENT FAILURE. THE RECOMMENDED INLET
SUPPLY PRESSURE IS
7” WATER COLUMN.
WHEN OPERATING THE ODORATOR 2 YOU WILL BE WORKING WITH FLAMMABLE
METHANE OR PROPANE GAS WHICH IS POTENTIALLY DANGEROUS IF NOT HANDLED
PROPERLY. METHANE GAS FROM A LECTURE BOTTLE MAY BE ODORLESS, COLORLESS
AND IS E X P L O S I V E IN CONCENTRATIONS OF APPROXIMATELY 5.0 TO 15.0 PERCENT BY
VOLUME IN
AIR. PROPANE’S EXPLOSIVE RANGE IS APPROXIMATELY 2.2 TO 9.6 PERCENT BY
VOLUME IN AIR.
PRIOR TO CONDUCTING TESTS, VERIFY THE ENVIRONMENT AND IMMEDIATE AREA DO NOT
CONTAIN ANY POSSIBLE IGNITION SOURCES.
WHEN NOT OPERATING THE ODORATOR 2, CLOSE ALL VALVES AND TURN THE ODORATOR 2
OFF. ANY TIME GAS IS BEING PASSED THROUGH THE ODORATOR 2 THE POWER MUST BE
TURNED ON. THIS WILL DILUTE THE GAS AT THE EXHAUST PORT AND PREVENT POCKETS
OF CONCENTRATED GAS FROM ACCUMULATING.
100% LEL IS APPROXIMATELY EQUAL TO 5.0% METHANE OR 2..2 % PROPANE GAS BY
VOLUME IN AIR. AS LONG AS A REFERENCE CGI READS SAFELY BELOW THE LEL (20% LEL
OR LESS) YOU SHOULD NOT HAVE ANY PROBLEMS WITH A FLAMMABLE MIXTURE
BUILDING UP.
ANY SUSPICION OF A GAS LEAK BY UNEXPLAINED SOUND OR OPERATION OF THE
ODORATOR 2 IS REASON TO IMMEDIATELY SHUT OFF THE GAS SUPPLY VALVES AND
THEN SEEK THE CAUSE.
IF INDOORS, DO NOT TURN THE ODORATOR 2 ON UNTIL A REFERENCE CGI INDICATES S A F E
Содержание Odorator 2
Страница 1: ...ODORATOR 2TM Operator s Manual All Models...
Страница 20: ...16 Periodic Test Records Test Date Results Action Operator Name...
Страница 21: ...17 Periodic Test Records Test Date Results Action Operator Name...
Страница 26: ...22...
Страница 28: ...Heath Consultants Incorporated 9030 Monroe Road Houston TX 77061 Tel 713 844 1300 www heathus com...