DO NOT operate unit if you are not in good health. People with a cardiac
condition should not operate this unit.
●
DO NOT operate this unit if you have a pacemaker. This unit should only be
used for checking porosity (or electrical breakdown) of dielectric or insulating
materials (e.g., Jacketing material, powder coatings).
●
DO NOT use this unit around other machinery. An electrical shock may cause
the operator to fall and injure themselves.
●
DO NOT operate this unit around people who not directly involved in the testing
procedure.
●
DO NOT operate this unit other than McGan Technology Probes. McGan
Probes CAN ONLY be used with the appropriate McGan Insulation Tester(s).
●
DO NOT simultaneously handle the brush electrode and ground clamp, as it will
cause a mild “tingle”. Use surgical gloves as a precaution against “tingle”.
●
You cannot operate the MM513 unit with the AC adaptor plugged into the
rechargeable battery port on the bottom of the base unit.
●
ONLY USE Lithium Polymer (LiPo) battery and the associated AC power
adapter (PN/5VQACP-0015) provided with the M513 system.
●
LED battery indicator light will illuminate when the unit is low on power.
o
If the power from the battery is too low the LED will not illuminate.
o
If the MM513 fails to operate due to battery failure, contact Customer
Service.
●
Always keep the working end of any of the probe electrodes away from your
body
●
Do not
touch probe electrodes when the instrument is activated.
●
After the instrument has been turned off, always ground the probe before
disassembling the unit to ensure that any residual charge has dissipated.
⚠
DANGER
⚠
●
DO NOT use wet.
After cleaning,
thoroughly
DRY all areas before using the
components,
and inspect for any defects in the electrodes.
●
DO NOT
use the test equipment in any combustible or flammable
atmosphere
(i.e., flammable anesthetics)
,
as a test voltage can cause an arc or
spark to be generated and
an explosion could result
.
Troubleshooting
Symptom
Possible Cause
Solution
No display
Dead or low
charged battery
Fully charge
battery pack.
Alarm sounds
continuously during test
Surface might be
slightly conductive,
damp, or salty.
Probe moved too
fast
Wash, clean, and
dry the surface.
Move probe
approximately 3
feet (1-m) every 4
seconds.
No alarm on defect
Voltage is too low.
Increase voltage
sensitivity.
No spark at probe tip
Damaged leads
Poor connections
Dead or low
charged battery
Repair or replace
leads.
Clean and
reconnect.
Recharge the
battery.
No battery indicator
light and unit does not
function
Dead or low
charged battery.
Recharge the
battery.
Disposal
N/A
Document Version
4