DO NOT operate unit if you are not in good health. People with a cardiac 
condition should not operate this unit.

DO NOT operate this unit if you have a pacemaker. This unit should only be 
used for checking porosity (or electrical breakdown) of dielectric or insulating 
materials (e.g., Jacketing material, powder coatings).

DO NOT use this unit around other machinery. An electrical shock may cause 
the operator to fall and injure themselves.

DO NOT operate this unit around people who not directly involved in the testing 
procedure.

DO NOT operate this unit other than McGan Technology Probes. McGan 
Probes CAN ONLY be used with the appropriate McGan Insulation Tester(s).

DO NOT simultaneously handle the brush electrode and ground clamp, as it will 
cause a mild “tingle”. Use surgical gloves as a precaution against “tingle”.

You cannot operate the MM513 unit with the AC adaptor plugged into the 
rechargeable battery port on the bottom of the base unit.

ONLY USE Lithium Polymer (LiPo) battery and the associated AC power 
adapter (PN/5VQACP-0015) provided with the M513 system.

LED battery indicator light will illuminate when the unit is low on power.

o

If the power from the battery is too low the LED will not illuminate.

o

If the MM513 fails to operate due to battery failure, contact Customer 
Service.

Always keep the working end of any of the probe electrodes away from your 
body

Do not

 touch probe electrodes when the instrument is activated.

After the instrument has been turned off, always ground the probe before 
disassembling the unit to ensure that any residual charge has dissipated.

 

DANGER

 

DO NOT use wet. 

After cleaning, 

thoroughly

 

DRY all areas before using the 

components,

 and inspect for any defects in the electrodes.

DO NOT

 

use the test equipment in any combustible or flammable 

atmosphere

 (i.e., flammable anesthetics)

,

 as a test voltage can cause an arc or 

spark to be generated and 

an explosion could result

.

 

Troubleshooting

Symptom

Possible Cause

Solution

No display

Dead or low 
charged battery

Fully charge 
battery pack.

Alarm sounds 
continuously during test

Surface might be 
slightly conductive, 
damp, or salty.

Probe moved too 
fast

Wash, clean, and 
dry the surface.

Move probe 
approximately 3 
feet (1-m) every 4 
seconds.

No alarm on defect

Voltage is too low.

Increase voltage 
sensitivity.

No spark at probe tip

Damaged leads

Poor connections

Dead or low 
charged battery

Repair or replace 
leads.

Clean and 
reconnect.

Recharge the 
battery.

No battery indicator 
light and unit does not 
function

Dead or low 
charged battery.

Recharge the 
battery.

Disposal

N/A

Document Version

4

Содержание MM513-100

Страница 1: ...rrent output 0 1 mA 0 0001 A at probe Short circuit test current 0 1 mA max Power supply 1800 mAh Li Polymer Battery Probes o Medical style brass wire eight 8 mm wide brush trim length of two 2 mm NOT...

Страница 2: ...Insulation Tester MM513 100 Bi Polar Fixture Black Connector MMBPT 190 Ring and Brush Electrodes MMLSE 0029 Ground Wire with Alligator Clip Tri Hole Electrode MMTRI 0022A Saddle Block MMSBT 170 Docum...

Страница 3: ...ons N A Packaging Conditions N A Shelf Life N A Instructions for Using Product Description of Use s For testing insulation integrity of electrosurgical instruments Preparation for Use 1 Turn the unit...

Страница 4: ...f the unit under test See Operational Guide for Saddle block Diagrams drawings pictures N A Steps for Use of Product 1 Remove the insulation tester unit and accessories from the carrying case 2 Take t...

Страница 5: ...pital s procedure policy with regards to the instrument under test Note The unit should always be switched off prior to removing or repositioning of the ground lead the HV red wire or the saddle block...

Страница 6: ...st Fig 8 Figure 8 4 Turn the base unit on and set the voltage to 2 8 0 3 kV Fig 9 Figure 9 5 Push the ESI under test through the LSE ring electrode slowly approximately three 3 feet every four 4 secon...

Страница 7: ...sert the end of one tine of the Bi Polar forceps into the middle of the brush Fig 11 Figure 11 5 Turn the base unit on and set the voltage to 2 8 0 3 kV 6 Slowly push the Bi Polar forceps away from yo...

Страница 8: ...side of tine c On the inside of the right tine d On the outside of the right tine Figure 15 9 If the alarm sounds the Bi Polar has a fault Follow standard facility procedures for a defective electrosu...

Страница 9: ...e LED will not illuminate o If the MM513 fails to operate due to battery failure contact Customer Service Always keep the working end of any of the probe electrodes away from your body Do not touch pr...

Страница 10: ...rly degradation Avoid contact with such materials It is recommended to calibrate the MM513 base unit P N MM513 110 at least once per year to ensure it is operating at the appropriate voltage Healthmar...

Страница 11: ...Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com Document Version 4...

Отзывы: