  Insulation Kit   

                                     

 Insulation Tester MM513-100

                                    

       

                               

(Bi-Polar Fixture Black Connector)-MMBPT-190

 

     

                

  

Ring and Brush Electrodes-MMLSE-0029              Ground Wire with Alligator Clip

     

  Tri-Hole Electrode- MMTRI-0022A                    Saddle Block- MMSBT-170      

Document Version

4

Содержание MM513-100

Страница 1: ...rrent output 0 1 mA 0 0001 A at probe Short circuit test current 0 1 mA max Power supply 1800 mAh Li Polymer Battery Probes o Medical style brass wire eight 8 mm wide brush trim length of two 2 mm NOT...

Страница 2: ...Insulation Tester MM513 100 Bi Polar Fixture Black Connector MMBPT 190 Ring and Brush Electrodes MMLSE 0029 Ground Wire with Alligator Clip Tri Hole Electrode MMTRI 0022A Saddle Block MMSBT 170 Docum...

Страница 3: ...ons N A Packaging Conditions N A Shelf Life N A Instructions for Using Product Description of Use s For testing insulation integrity of electrosurgical instruments Preparation for Use 1 Turn the unit...

Страница 4: ...f the unit under test See Operational Guide for Saddle block Diagrams drawings pictures N A Steps for Use of Product 1 Remove the insulation tester unit and accessories from the carrying case 2 Take t...

Страница 5: ...pital s procedure policy with regards to the instrument under test Note The unit should always be switched off prior to removing or repositioning of the ground lead the HV red wire or the saddle block...

Страница 6: ...st Fig 8 Figure 8 4 Turn the base unit on and set the voltage to 2 8 0 3 kV Fig 9 Figure 9 5 Push the ESI under test through the LSE ring electrode slowly approximately three 3 feet every four 4 secon...

Страница 7: ...sert the end of one tine of the Bi Polar forceps into the middle of the brush Fig 11 Figure 11 5 Turn the base unit on and set the voltage to 2 8 0 3 kV 6 Slowly push the Bi Polar forceps away from yo...

Страница 8: ...side of tine c On the inside of the right tine d On the outside of the right tine Figure 15 9 If the alarm sounds the Bi Polar has a fault Follow standard facility procedures for a defective electrosu...

Страница 9: ...e LED will not illuminate o If the MM513 fails to operate due to battery failure contact Customer Service Always keep the working end of any of the probe electrodes away from your body Do not touch pr...

Страница 10: ...rly degradation Avoid contact with such materials It is recommended to calibrate the MM513 base unit P N MM513 110 at least once per year to ensure it is operating at the appropriate voltage Healthmar...

Страница 11: ...Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com Document Version 4...