Healthmark MM513-100 Скачать руководство пользователя страница 4

a. If the unit’s battery level is red, recharge using the charger adapter supplied 

with the insulation tester kit. (NOTE: Use of any other charger may cause 
damage to the insulation tester unit and void warranty.)

b. Only use this unit if it is at Green=Battery Full.

3. Connect the HV probe and ground leads to the unit.
4. Connect the ground clamp to the metallic substrate of the item to be tested. 

Substrate should be grounded.

5. Attach the selected probe (ring, brush, Tri-Hole, or Saddle block) to the HV 

Wire or base unit port (red).

6. Power on unit and select voltage.
7. Place the probe near the metal substrate.
8. A spark should occur (if not recheck all leads).
9. The unit should now be ready for use.
10. Test the coated surface by lightly moving the probe (brush, ring, and Tri-Hole 

electrode) slowly (approximately three [3] feet every four [4] seconds) across 
the surface of the unit under test. (See Operational Guide for Saddle block.)

Diagrams (drawings, pictures)

N/A

Steps for Use of Product

1. Remove the insulation tester unit and accessories from the carrying case.
2. Take the green ground wire and firmly insert it into the green port on the bottom 

of the base unit. 

(Fig. 1)

.

Figure 1 Insulation Tester

3. Secure the Saddle block to a flat (preferably metal) surface by pushing down on 

the top of the unit until the suction feet stick to the surface. 

(Fig. 2)

.

Figure 2

4. Attach the red port on the top of the insulation tester unit directly to the side pin 

of the Saddle block. Make sure controls face up.  

(Fig. 3)

.

        

              

             

 Figure 3 

                                                      

 Figure 3

There are several ways to set the Saddle block up depending upon the 
electrosurgical instrument (ESI) to be tested and/or the kit used.

1. Insert the chosen electrode securely into the proper slot on the Saddle block.
2. Take the clamp on the green ground wire and attach it to the conductive core of 

the instrument under test.

3. Turn the base unit on and set the voltage to 2.8- ± 0.3 kV. 

(Fig. 4)

.

Document Version

4

Содержание MM513-100

Страница 1: ...rrent output 0 1 mA 0 0001 A at probe Short circuit test current 0 1 mA max Power supply 1800 mAh Li Polymer Battery Probes o Medical style brass wire eight 8 mm wide brush trim length of two 2 mm NOT...

Страница 2: ...Insulation Tester MM513 100 Bi Polar Fixture Black Connector MMBPT 190 Ring and Brush Electrodes MMLSE 0029 Ground Wire with Alligator Clip Tri Hole Electrode MMTRI 0022A Saddle Block MMSBT 170 Docum...

Страница 3: ...ons N A Packaging Conditions N A Shelf Life N A Instructions for Using Product Description of Use s For testing insulation integrity of electrosurgical instruments Preparation for Use 1 Turn the unit...

Страница 4: ...f the unit under test See Operational Guide for Saddle block Diagrams drawings pictures N A Steps for Use of Product 1 Remove the insulation tester unit and accessories from the carrying case 2 Take t...

Страница 5: ...pital s procedure policy with regards to the instrument under test Note The unit should always be switched off prior to removing or repositioning of the ground lead the HV red wire or the saddle block...

Страница 6: ...st Fig 8 Figure 8 4 Turn the base unit on and set the voltage to 2 8 0 3 kV Fig 9 Figure 9 5 Push the ESI under test through the LSE ring electrode slowly approximately three 3 feet every four 4 secon...

Страница 7: ...sert the end of one tine of the Bi Polar forceps into the middle of the brush Fig 11 Figure 11 5 Turn the base unit on and set the voltage to 2 8 0 3 kV 6 Slowly push the Bi Polar forceps away from yo...

Страница 8: ...side of tine c On the inside of the right tine d On the outside of the right tine Figure 15 9 If the alarm sounds the Bi Polar has a fault Follow standard facility procedures for a defective electrosu...

Страница 9: ...e LED will not illuminate o If the MM513 fails to operate due to battery failure contact Customer Service Always keep the working end of any of the probe electrodes away from your body Do not touch pr...

Страница 10: ...rly degradation Avoid contact with such materials It is recommended to calibrate the MM513 base unit P N MM513 110 at least once per year to ensure it is operating at the appropriate voltage Healthmar...

Страница 11: ...Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com Document Version 4...

Отзывы: