Instructions for Use: Insulation Tester and Bi-Polar Fixture

Brand Name of Product

Insulation Tester and Bi-Polar Fixture

Generic Name of Product

Insulation Tester and Bi-Polar Fixture

Product Code Number(s)

MM513-100, MM513-600, MM513-120EU, MMBPT-190, MMBRU-007, MMLSE-
0029, MMTRI-0022A, MMSBT-170

Intended Use

Compact, handheld, battery operated unit that tests the integrity (e.g., pinholes, cracks, or 
defects) of the insulation of electrosurgical instruments to prevent tissue burns of 
laparoscopic and bipolar electrosurgical instruments.

Range of Applications for Product

Used to test the integrity of the insulation of electrosurgical instruments (ESI) and 
cable/cords for laparoscopic, endoscopic, intraoperative instruments, monopolar and 
bipolar surgical items.

Key Specifications of Product

Insulation Tester- MM513-100

Bi-Polar Fixture and Black Connector—MMBPT-190

Training CD

Quick Operation Guide (manual)

On and off switch

Lightweight portable unit

Measurements:

Weight: 0.672 pounds (305 g)

Voltage: 0- to 5 kV fully adjustable

Resolution: 10 V

Current output: < 0.1 mA (0.0001 A) at probe.

Short circuit: test current < 0.1 mA max.

Power supply: 1800 mAh Li-Polymer Battery.

Probes:

o

Medical style brass wire eight (8) mm wide brush, trim length of two 
(2) mm. (NOTE: Probe size/shape may vary depending on user 
requirement)

o

Reusable/interchangeable brush, ring, or Tri-Hole electrode

o

Test stand—Saddle Block.

System case

Ground wire (green wire):

with alligator clip—six (6) feet

with clamp on one end—two (2) m.

Power supply option: 5 V external rechargeable battery with adaptor.

Lithium Polymer battery.

Dimensions: 8.5- x 3.1- x 1.5 inches (215- x 78- x 38 mm)

Ten (10)-hour operational time (up to 1000 instruments), two (2)- to 4-hour 
recharging cycle.

Simple operation, easy to read LED indicators.

Maintains applied test voltage with constant current source.

Full test current at low voltages.

Limited output current for operational safety.

Voltage up and down buttons

LED displays alarm (audible and visual) and battery charge

C-Style adapter wall charger

Brush Range Voltage

Brush 

Electrode

MMBRU-007

LS Ring 

Electrode

MMLSE-0029

Tri-Hole 

Electrode

MMTRI-

0022A

Bi-Polar 

Fixture

MMBPT-

190

Wire Tester 

MMWIT-

200A

3.0 ± 0.3 kV

2.8 ± 0.3 kV

4.2 ± 0.3 kV

2.8 ± 0.3 kV

4.2 ± 0.3 kV

Kit Contents

Insulation Kit Includes:

Document Version

4

Содержание MM513-100

Страница 1: ...rrent output 0 1 mA 0 0001 A at probe Short circuit test current 0 1 mA max Power supply 1800 mAh Li Polymer Battery Probes o Medical style brass wire eight 8 mm wide brush trim length of two 2 mm NOT...

Страница 2: ...Insulation Tester MM513 100 Bi Polar Fixture Black Connector MMBPT 190 Ring and Brush Electrodes MMLSE 0029 Ground Wire with Alligator Clip Tri Hole Electrode MMTRI 0022A Saddle Block MMSBT 170 Docum...

Страница 3: ...ons N A Packaging Conditions N A Shelf Life N A Instructions for Using Product Description of Use s For testing insulation integrity of electrosurgical instruments Preparation for Use 1 Turn the unit...

Страница 4: ...f the unit under test See Operational Guide for Saddle block Diagrams drawings pictures N A Steps for Use of Product 1 Remove the insulation tester unit and accessories from the carrying case 2 Take t...

Страница 5: ...pital s procedure policy with regards to the instrument under test Note The unit should always be switched off prior to removing or repositioning of the ground lead the HV red wire or the saddle block...

Страница 6: ...st Fig 8 Figure 8 4 Turn the base unit on and set the voltage to 2 8 0 3 kV Fig 9 Figure 9 5 Push the ESI under test through the LSE ring electrode slowly approximately three 3 feet every four 4 secon...

Страница 7: ...sert the end of one tine of the Bi Polar forceps into the middle of the brush Fig 11 Figure 11 5 Turn the base unit on and set the voltage to 2 8 0 3 kV 6 Slowly push the Bi Polar forceps away from yo...

Страница 8: ...side of tine c On the inside of the right tine d On the outside of the right tine Figure 15 9 If the alarm sounds the Bi Polar has a fault Follow standard facility procedures for a defective electrosu...

Страница 9: ...e LED will not illuminate o If the MM513 fails to operate due to battery failure contact Customer Service Always keep the working end of any of the probe electrodes away from your body Do not touch pr...

Страница 10: ...rly degradation Avoid contact with such materials It is recommended to calibrate the MM513 base unit P N MM513 110 at least once per year to ensure it is operating at the appropriate voltage Healthmar...

Страница 11: ...Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com Document Version 4...