Page
20
of
24
4/24/2018
* Please contact Healthmark Industries for a Return Materials Authorization
to have the scope evaluated.
Advanced Settings
at surface.
Low Light
No image or very
dark image.
Weak or light pattern
when scope pointed
at surface.
Broken light fibers
in scope.
Replace Flexible
Inspection Scope.*
(The user must judge
when the Flexible
Inspection Scope is no
longer adequate for
use, but as a guide,
when 10% of the
image or illumination
has been degraded or
lost it is recommended
that a replacement be
acquired).
No image or
distorted
image
No image or heavily
distorted, cracked
appearance.
Broken image
sensor and/or
internal cables.
Replace Flexible
Inspection Scope.*
Overly bright
image
White-out type
reflection.
Light Intensity
is
too bright.
Press Decrease (-)
Brightness Button to
decrease light output
Blurry Image
or overly
bright image
Distorted image.
Light often reflects
back and image
appears brightly
colored.
Debris or film on
lens.
Wipe off end of
Flexible Inspection
Scope with soft cloth.
Image does
not capture
When you click the
Capture Button, the
still image or video is
not captured.
The File Loc path
may have changed
or the folder name
does not exist.
Set up a new
Windows File Loc
folder.
Rapidly
takes
pictures
automatically
‘Picture Captured’
keeps flashing and
Image files are
created rapidly.
PC’s internal
camera is selected
as the video
device in Settings.
Disable the PC’s
internal camera (see
Advanced Settings
below).