2. HMDI is “ON”.

Low light 

No image or very 
dark image. Weak 
light pattern when 
scope pointed at 
surface.

Light setting too 
low.

Cycle through light 
intensity levels/settings 
until a clear image is 
obtained.

Low light 

No image or very 
dark image. Weak 
or light pattern 
when scope 
pointed at surface.

Broken light 
fibers in scope.

 Replace Flexible 
Inspection Scope™

Decide if the 
Scope is no 
longer adequate 
for use.

Recommendation 
is when 10% of 
the image or 
illumination has 
been degraded/ 
lost to replace 
the scope. 

No image or 
distorted image 

No image or 
heavily distorted, 
cracked 
appearance. 

Broken image 
sensor and/or 
internal cables.

1.Press Power Cycle 
button
2.Replace Flexible 
Inspection Scope™.

Overly bright 
image

White-out type 
reflection

Light intensity is 
too bright

Cycle through light 
intensity levels/settings 
until a clear image is 
obtained. 

Blurry image or 
overly bright 
image

Distorted image. 
Light often 
reflective and 
image appears 
brightly colored.

Debris or film on 
lens.

Wipe off end of Flexible 
Inspection Scope™ with 
non-Linting wipe.

Image does not 
capture

When you click 
the Capture 
Button, the still 
image or video is 
not captured.

The File Loc path 
may have 
changed, or the 
folder names do 
not exist.

Set up a new Windows 
File Loc folder.

Rapidly takes 
pictures 
automatically

‘Pictured 
Captured’ keeps 
flashing and 
Image files are 
created rapidly.

PC’s internal 
camera is 
selected as the 
video device is 
Settings.

Disable the PC’s internal 
camera (See Advanced 
Settings below).

Reassembly Instructions

N/A

Packaging

N/A

Sterilization

Do Not autoclave

 the Flexible Inspection Scope™. 

See the Chemical Compatibility Chart (PDF): 

Click here.

Low Temperature Sterilization Systems

 

(Flexible Inspection Scope Only)

Ethylene Oxide (EtO)

STERRAD

®

 100S System (Standard)

STERRAD

® 

NX System (Standard, 

Advanced)

STERRAD

®

 100NX System (Standard)

STERIS

®

 Liquid Chemical Sterilization 

Systems

STERIS V-PRO

®

 Low Temperature 

Sterilization Systems (Non-Lumen Cycle)

Storage

Storage and transport

o

Humidity: 10 to 100% rh

o

Temperature: 

−20-

 to 60 °C

Pressure: 600- to 900 hPa

o

Normal Operation

o

Humidity: Zero (0) to 100% rh

o

Temperature: 5- to 40 °C

Document Version

2

Содержание CT-101

Страница 1: ...ls o Field of View 120 in air o Angle of view 0 HDMI Control Module 2 mm Control module housing Camera processor and LED illumination HDMI Dimensions 5 24 x 3 90 x 1 85 inches Weight 1 20 pounds Digital Inspection Scope Connection Illumination Control Power Cycle Power Input 100 to 240 volts V alternating current ac 1 ampere A 47 to 63 hertz Hz from wall outlet Power Output 5 V 2 A to enclosure Sy...

Страница 2: ...g Product Description of Use s For visually inspecting items Preparation for Use Unpacking Flexible Inspection Scope Carefully inspect for shipping damage If there is any damage contact the shipping carrier and Heatlhmark customer service 800 521 6224 immediately HDMI Control Module Fig 1 1 Digital Inspection Scope Connection 2 Illumination Control 3 Power Cycle 4 N A 5 Mini HDMI Out 6 Power Input...

Страница 3: ...gure 3 Power Cycle Button Press button to RESET camera Fig 4 Figure 4 1 Flexible Inspection Scope Plug Fig 5 Contains camera video connection as well as LED Light for illumination Figure 5 2 Flexible Working Length Fig 6 The portion of the Flexible Inspection Scope that is inserted into an item during visual inspection The measuring scale markings on the Flexible Working Length are in centimeters ...

Страница 4: ...Different Flexible Inspection Scope on the Control Module 1 Press the Power button on the Control Module 2 Disconnect the current Flexible Inspection Scope from the Control Module 3 Repeat the steps in the STARTING SOFTWARE CONNECTIONS SCOPE TO PC procedure Diagrams drawings pictures Inserting Scope in Item 2 3 Document Version 2 ...

Страница 5: ...rol box that is in use then disconnect the Flexible Inspection Scope from the control box 6 If the HDMI control box is in use power off the control box disconnect the Borescope Catheder and along with the power adapter NOTE If unable to exchange catheters by recycling power close the program and open again Interpretation of Test Results N A Contraindications of Test Results N A Documentation N A S...

Страница 6: ...lcohol Ethoxylates Neutral or Near Neutral pH Detergents Enzymatic Cleaning Solutions Enzymatic Detergents Glutaraldehyde Hydrogen Peroxide 7 35 with Peracetic Acid 0 23 Accelerated Hydrogen Peroxide AHP Ortho phthalaldehyde OPA Peracetic Acid Sodium Borated Decahydrate Sodium Xylene Sulfonate Tetrapotassium Pyrophosphate Trisodium nitrilotriacetate The Flexible Inspection Scope has fluid ingress ...

Страница 7: ...g Prior to use carefully inspect the external surfaces of the Flexible Inspection Scope and any accessories to ensure they are smooth and free of any wear or damage i e protrusions or sharp edges Flexible Inspection Scope have no user o serviceable parts o maintenance beyond cleaning Refer all service or replacement needs to Healthmark Industries Light leaks may be common and possibly noticeable w...

Страница 8: ...or film on lens Wipe off end of Flexible Inspection Scope with non Linting wipe Image does not capture When you click the Capture Button the still image or video is not captured The File Loc path may have changed or the folder names do not exist Set up a new Windows File Loc folder Rapidly takes pictures automatically Pictured Captured keeps flashing and Image files are created rapidly PC s intern...

Страница 9: ...quency for cleaning disinfection and sterilization This assessment should be based upon clinical use of items and reprocessing instructions Related Healthmark Products N A Other Product Support Documents ProSys Brochure ProSys Price List Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com ...