HARDWARE INSTALLATION -
Step 2 for Connex
®
Monitors
The following instructions illustrate the hardware set up for connecting to a Welch Allyn
®
Connex
®
Vital Signs Monitor (CVSM). USB ports on the Connex Spot and Integrated Wall Systems are
located on the underside of the monitor.
1.
Connect the USB Wireless Dongle to the HOMWA unit as the following figure.
2.
Connect the HOMWA (black box) with dongle to the Welch Allyn
®
CVSM as shown below.
3.
Power on the Welch Allyn
®
CVSM unit and power on the scale to initialize wireless
communication. The connection is now established.
4.
To enable weight scale communication on the Welch Allyn
®
CVSM, follow these steps.
a) Connect the CVSM to a PC to access the Welch Allyn
®
Service Tool. This service tool either
comes with the Welch Allyn
®
device or can be downloaded at
www.welchallyn.com/en/service-support/service-center/service-tool.html.
b) Follow the prompts in the Service Tool to activate the weight scale license.
c) Activate the license by entering the authorization code A66FF29A3B2F85E1
Continue to page 12 for Step 3 - Wireless Communication instructions.
Note:
The set up presumes that the HOMBT1 Scale Unit is attached to the scale, see Step 1 on
page 5.
HOMWA Unit Connection to
CVSM
1) Connect the 1 ft. USB cable to
any available USB port on the
Welch Allyn
®
CVSM
2) Connect the other end of the
1 ft. USB cable to the “
TO
CONNEX DEVICE
” USB port
on the HOMWA (black box)
Welch Allyn
®
CVSM
Connect the
USB Wireless
Dongle
to the “
TO SCALE
” USB
Port on the HOMWA unit
HOMWA Unit
(Black Box
with White Label)