CAUTIONS AND WARNINGS 

Intended use 

The Health o meter

®

 Professional 524KL / 524KG scale is intended to be used in a professional medical 

environment by trained medical staff. This product was designed to weigh infants that are sitting in the weighing 
seat. Do not modify the product or use it for anything other than its intended purpose. 
 

To prevent injury and damage to your scale, please follow these instructions very carefully. 

  Assemble and operate the scale per the enclosed user instructions. 

  For accurate weighing, this scale must be placed on a flat, stable surface. 

  For accurate weighing, verify before each use the proper operation according to the procedure 

described in this manual. 

  Do not transport the scale with a patient or object on the seat. 

  Do not exceed the weight capacity specified for this scale.  

  Do not use in the presence of flammable or explosive materials. 

  Ensure that neither the scale nor the AC adapter come in contact with liquids, excessive 

temperature, or excessive humidity. 

  If the scale becomes damaged, it should not be operated until properly serviced.  

 

Patient/caregiver safety 

  This scale is designed for static weighing of patients only. No scale should be used for patient 

transfer.  

  To prevent patient injury, the patient must be attended throughout the entire weighing event.   

 

When operating the scale with batteries: 

  I

f the “LO” indicator activates, replace the batteries or switch to the AC adapter as soon as possible.  

  When replacing / inserting batteries, be sure to use all new batteries. 

 
When operating the scale with the (optional) AC adapter: 

  Operate this scale exclusively with the AC adapter that is specified (ADPT31). Use of an unspecified 

adapter will void the warranty, and can pose a serious safety hazard.   

  Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent plug 

prongs. 

  Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage 

appropriate for operation.   

  Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.  

  Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the 

scale.   

  Operating this scale at voltages and frequencies other than specified can damage the equipment 

and will void the warranty. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 
93/42/EEC (EN 55011 Class A and EN 60601-1-2). These limits are designed to provide reasonable 
protection against harmful interference in a typical medical installation. The equipment generates, uses 
and may radiate radio frequency energy and, if not installed and used in accordance with these 
instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that 
interference will not occur in a particular installation. If this equipment does cause interference with other 
devices, which can be determined by turning the equipment off and on, the user is encouraged to try to 
eliminate the interference by one or more of the following measures: 

o

 

Reorient or relocate the receiving device.  

o

 

Increase the separation between the equipment. 

o

 

Plug the equipment into an outlet on a circuit different from that to which the other device(s) is 
connected. 

o

 

Consult Customer Service or a field service technician for help. 

 

In no event whatsoever shall Pelstar, LLC be liable for damages or injuries arising 

from or connected with the assembly, use, or misuse of its products. 

Содержание 524KG

Страница 1: ...1 Model 524KL 524KG Digital Pediatric Seat Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM524KL_524KG Rev20180227...

Страница 2: ...l 5 Assembly Instructions 6 Powering the Scale 9 Set Up Instructions 10 Operating Instructions 11 Maintenance 12 Calibration 13 Troubleshooting 14 Warranty 15 Note This scale has been factory calibrat...

Страница 3: ...e of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Prior...

Страница 4: ...0 20 lb 0 2 oz 20 50 lb 0 5 oz 0 9 kg 5g 9 23 kg 10 g 524KG 23 kg 0 9 kg 5g 9 23 kg 10 g Power Requirements 6 AA batteries included optional 120V power adapter ADPT31 Environmental Operating temperat...

Страница 5: ...tivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please contac...

Страница 6: ...ind inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton asid...

Страница 7: ...column and pull gently until the plug reaches the bottom of the column Carefully pull the cable connector from the bottom of the column and plug into the RJ jack on the scale platform Caution Once con...

Страница 8: ...brackets and into the platform Using a Phillips head screwdriver tighten the screws to secure both brackets to the platform IMPORTANT Ensure that bracket C is mounted on the edge of the platform not t...

Страница 9: ...Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measu...

Страница 10: ...te in the selected mode 5 Repeat steps 3 4 to change the weighing mode to a different unit of measure LB KG Lock Out The weight measuring unit can be locked to only measure in KG or LB KG Lock Instruc...

Страница 11: ...e HOLD RELEASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the scale 6 Press the ON OFF button to turn the...

Страница 12: ...r to prevent battery depletion do not store or leave anything on the scale Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Caution Disconnect t...

Страница 13: ...ating that you can begin calibration Release all three buttons CAL followed by XXXX Press the ZERO button C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four...

Страница 14: ...eighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not sitting still Try to have the patient sit still The display shows Over messa...

Страница 15: ...d tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To ge...