MAINTENANCE 

Maintenance 

The following pages provide instructions for maintenance, cleaning, calibrating, and 
troubleshooting of your scale. Maintenance operations other than those described in this 
manual should be performed by qualified service personnel. 

Caution: Before first use, or after long periods of non-use, check the scale for proper 
operation and function. If the scale does not operate correctly, refer to qualified service 
personnel. 

1. Check overall appearance of the total scale for any obvious damage, wear, and tear. 
2. Inspect the AC adapter cord for cracking or fraying, or for broken/bent prongs. 
 

Note:

 This scale is an extremely sensitive weighing device. Items such as clothing or towels 

placed or dropped on the scale may cause the scale to activate. In order to prevent battery 
depletion, do not store or leave anything on the scale.  
 

Cleaning 

Proper care and cleaning is essential to ensure a long life of accurate and effective operation.  

Caution: Disconnect the scale from the AC adapter power source prior to cleaning the 
unit.   

1. Clean all external surfaces with a clean damp cloth. Mild antimicrobial soap and water 
solution may be used. Dry with a clean soft cloth. 
2. Do not immerse the scale into cleaning or other liquid solutions or let cleaning solutions pool 
on the scale’s surfaces. 
3. Do not use Isopropyl Alcohol or other solutions to clean the display surface.  
4. Do not use abrasive cleaners. 
  

Replacement Parts 

Parts may be available if a piece of the scale needs to be replaced. Contact Health o meter

®

 

Professional Scales Customer Service at 1-800-815-6615 to inquire about the availability of these 
replacement parts. 

Part # 

Description 

522KL/524KLKEYPAD 

Keypad for 524KL 

524SEAT 

Seat for 524KL / 524KG 

ADPT31 

Power adapter 

Содержание 524KG

Страница 1: ...1 Model 524KL 524KG Digital Pediatric Seat Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM524KL_524KG Rev20180227...

Страница 2: ...l 5 Assembly Instructions 6 Powering the Scale 9 Set Up Instructions 10 Operating Instructions 11 Maintenance 12 Calibration 13 Troubleshooting 14 Warranty 15 Note This scale has been factory calibrat...

Страница 3: ...e of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Prior...

Страница 4: ...0 20 lb 0 2 oz 20 50 lb 0 5 oz 0 9 kg 5g 9 23 kg 10 g 524KG 23 kg 0 9 kg 5g 9 23 kg 10 g Power Requirements 6 AA batteries included optional 120V power adapter ADPT31 Environmental Operating temperat...

Страница 5: ...tivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please contac...

Страница 6: ...ind inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton asid...

Страница 7: ...column and pull gently until the plug reaches the bottom of the column Carefully pull the cable connector from the bottom of the column and plug into the RJ jack on the scale platform Caution Once con...

Страница 8: ...brackets and into the platform Using a Phillips head screwdriver tighten the screws to secure both brackets to the platform IMPORTANT Ensure that bracket C is mounted on the edge of the platform not t...

Страница 9: ...Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measu...

Страница 10: ...te in the selected mode 5 Repeat steps 3 4 to change the weighing mode to a different unit of measure LB KG Lock Out The weight measuring unit can be locked to only measure in KG or LB KG Lock Instruc...

Страница 11: ...e HOLD RELEASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the scale 6 Press the ON OFF button to turn the...

Страница 12: ...r to prevent battery depletion do not store or leave anything on the scale Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Caution Disconnect t...

Страница 13: ...ating that you can begin calibration Release all three buttons CAL followed by XXXX Press the ZERO button C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four...

Страница 14: ...eighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not sitting still Try to have the patient sit still The display shows Over messa...

Страница 15: ...d tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To ge...

Отзывы: