POWERING THE SCALE

 

Powering the Scale 

– Inserting Batteries 

1. With a Phillips head screwdriver remove the screw from the battery compartment cover 
located on the back of the display. Remove the battery cover by pressing downward on the 
locking tab. 
 
 
 
 
 
 
 
 
 
 
2. Replace or install 6 new AA batteries. Health o meter

®

 Professional recommends using an 

alkaline or lithium battery, rated for 130°F (54°C) operation.  Do not use zinc-carbon batteries. 
3. Re-attach the battery cover to the display assembly. 
4. Ensure all objects are removed from the weighing tray, and then press the 

ON/OFF

 button to 

turn the scale on. 

“88:8:8.8” followed by “0:0.0” will appear on the display.  

Note:

 The battery icon on the display does not indicate battery life. The battery icon simply 

indicates that batteries are being used as the power source. 

Note

: If the set up procedure fails, refer to the troubleshooting instructions. If the problem is not 

corrected, contact Health o meter Professional Scales Customer Service at 1-800-815-6615. 

Note:

 

If “Lo” appears on the display, it indicates that the battery power is low. It does NOT 

indicate a problem with the weight measurement. 

Warning: If the scale will not be used for some time, remove the batteries to avoid a 
safety hazard. Disposal of batteries must be performed in accordance with the prevailing 
national, regional or local regulations which apply to you

.

   

 
Optional AC Adapter 

1.  Connect the AC adapter (not included) into the adapter jack on the scale platform. 

Warning: To avoid a safety hazard, use only a Health o meter

®

 Professional AC adapter. 

 

 
 
 
 
 
 
 

 

2. Plug the 

scale’s AC adapter into the power source/wall outlet. 

Caution:

 Always plug the AC 

adapter into the scale first, then into the power source/wall outlet. 
3. When the AC adapter is disconnected from the scale, the scale will switch to battery power if 
batteries are installed.  

Note

: If the set up procedure fails, refer to the troubleshooting instructions. If the problem is not 

corrected, refer to qualified service personnel. 

 

 

Locking tab 

Battery cover 

Display assembly 

Содержание 522KG

Страница 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Страница 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Страница 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Страница 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Страница 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Страница 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Страница 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Страница 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Страница 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Страница 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Страница 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Страница 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Страница 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Страница 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Страница 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...

Отзывы: