ASSEMBLY INSTRUCTIONS (CONTINUED) 

 

5.  Position and secure the 

column to the scale platform. 
Locate the 4 column mounting 
screws previously removed in 
step 1. Insert the 4 screws 
through the bottom of the 
scale platform screw holes and 
into the column. Ensure that 
washers are located between 
the screw heads and the scale 
platform screw holes. Using a 
Phillips head screwdriver, 
tighten the screws to secure 
the column to the base of the 
platform. 

 
 
 
 
 
 
 
 
 
 
 

6.  Locate the 2 M6 x 16mm 

screws included in the clear 
plastic bag. Align the 2 holes 
in the tray with the mounting 
holes on the scale platform. 
Insert the screws and tighten 
to attach the tray to the scale 
platform.  

 
 
 
 
 
 
 
 

7.  Obtain the circular adhesive 

screw covers and place onto 
the screw cap on the tray.  

 
 
 
 
 

 

8.  Obtain the measuring tape and 

remove the adhesive backing. 
Affix the measuring tape to the 
bottom or side of tray.  

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Column 

(top) 

 

Column 

(top) 

 

Содержание 522KG

Страница 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Страница 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Страница 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Страница 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Страница 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Страница 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Страница 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Страница 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Страница 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Страница 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Страница 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Страница 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Страница 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Страница 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Страница 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...