25
Z6AM
HBM
A1559-2.0 de/en/fr
9
Specifications for the Z6... load cells
Type
Z6FC4
Z6FC6
Accuracy class according to OIML R 60
Maximal numbers of load cell verification
intervals (n
LC
)
C4
4000
C6
6000
Max. capacity (E
max
)
kg
20; 50; 100; 200;
500
50; 100; 200;
Minimum load cell verification interval
(v
min
)
% of
E
max
0.0066
Sensitivity (C
n
)
mV/V
2
Tolerance on sensitivity or load application
in specified direction
%
0,05
Temperature effect on sensitivity (TK
C
)
1)
% of
C
n
/10 K
0.0070
0.0040
Temperature effect on zero balance (TK
0
)
C
n
/10 K
0.0093
0.0093
Hysteresis error (d
hy
)
1)
0.0130
0.0080
Linearity deviation (d
lin
)
1)
% of C
n
0.0150
0.0110
Creep (d
DR
) over 30 min.
n
0.0125
0.0083
Input resistance (R
LC
)
Ω
350...480
Output resistance (R
0
)
Ω
356
0.12
Reference voltage (U
ref
)
V
5
Nominal range of excitation voltage (B
u
)
V
0.5...12
Insulation resistance (R
is
)
G
Ω
>5
Nominal temperature range (B
T
)
−10...+40 [+14...+104]
Service temperature range (B
tu
)
o
C [
o
F]
−30...+70 [−22...+158]
Storage temperature range (B
tl
)
C [ F]
−50...+85 [−58...185]
Safe load limit (E
L
)
% of
E
150
Breaking load (E
d
)
% of
E
max
300
Nennlast
kg
100
200
500
1000
Permissible dynamic load
% of
E
max
100
100
70
100
Nominal (rated) displacement, approx.
mm
0.31
0.39
0.6
0.55
Weight, approx.
kg
0.5
0.5
0.5
2.3
Degree of protection according to
EN60529 (IEC529)
IP 68
( severe test conditions: 1 m water
column ; 100 h )
Material
Measuring body
Bellows
Screwed cable gland
Cable sheath
Stainless steel
Stainless steel
Stainless steel / Viton
PVC
1)
The values for linearity deviation, relative reversibility error and temperature effect on sensitivity are typical
values. The sum of these values is within the cumulative error limits laid down by OIML R60.
Содержание Z6AM
Страница 2: ...Deutsch Seite 3 14 English Page 15 26 Fran ais Page 27 38...
Страница 39: ......