WE2111
A3983-1.0 en/de/fr
31
"
Press the
button.
CAP1
is displayed (capacity range 1).
"
Press the
button.
The factory setting displays
0030.00
, in all other
cases, the last input figure is displayed.
"
Press the
button until the 3 flashes.
"
Press the
button until a 1 appears instead of the 3.
"
Press the
button.
The maximum capacity of the scale is now 10.00.
CAP1
is displayed.
"
Press the
button.
E1
is displayed (minimum verification interval range
1). The default setting for the minimum verification
interval is 1 and does not therefore need to be
changed in this example.
"
Press the
button until
UnitS
is displayed.
"
Press the
button.
"
If
kg
is not selected as the unit next to the display,
press the
button until the unit is displayed.
"
Press the
button.
UnitS
is displayed.
"
Press the
button for several seconds to save the
settings.
The following are displayed in sequence before the
actual measured value appears:
SAviNG
, firmware
Содержание we2111