Mounting instructions
1: PW15AHi mounting at load application: 2 x M6; property class 10.9;
tightening torque 10 N
m; max. screw‐in depth 10 mm.
recommended flatness 0.1 mm on the connecting plate
2: PW15AHi mounting with mounting base: 2 x M6; property class 10.9:
tightening torque 10 N
m; max. screw‐in depth 10 mm
Mounting base
Recommended
flatness: 0.1
Platform center
Load direction
Platform
Load
application
(customer
design)
(1)
(2)
Fig. 4.2
PW15AHi mounting instructions
The PW15AHi load cell must be mounted on a clean surface with a flatness
better than 0.1 mm. Overload protection can be integrated in the form of a
M6x0.5 fine‐thread screw.
It is recommended that the load is applied in the center of the platform, to
minimize off‐center load errors and moments.
Before installing
several
PW15AHi into an installation with a bus system, take
the following into account:
The printed production number (typeplate) is required for setting up data
communication. If the typeplate can no longer be seen after installation, the
numbers of each PW15AHi should be noted beforehand. This enables different
addresses to be assigned during the initial start‐up.
Alternatively,
before
connection to the bus system, you can connect each
PW15AHi individually with a PC, in order to set different addresses (see ADR
command, Online Help).