15
PW15AHi
A2526-4.0 en/de/fr
HBM
6
Electrical connection
6.1
PW15AHi pin assignments
IMPORTANT
The PW15AHi must be operated with a supply voltage up to 30 V.
Incorrect connections between the supply and the interface cables can
cause irreversible damage.
Check the correct assignment of the connections before switching on for
the first time.
Any HBM guarantee is rendered invalid for damage caused by incorrect
connections.
The PW15AHi is fitted with an 8 pin connector socket.
Connector socket,
8 pin, made by
Lumberg
8
2
3
4
6
7
1
5
Fig. 6.1
Electrical connection
Terminal RS‐485
CANopen
DeviceNet
Color code for
1‐Kab165‐3
1)
1
GND
GND
GND
white
2
unassigned
unassigned
unassigned
brown
3
RA
CanH in
CanH in
green
4
Trigger / Stop Dosing
Trigger / Stop Dosing
Trigger / Stop Dosing
yellow
5
RB
CanL in
CanL in
gray
6
TB
CanL out
CanL out
pink
7
TA
CanH out
CanH out
blue
8
12...30 V
12...30 V
12...30 V
red
1)
Cable 1‐KAB165‐3 is available as an accessory.
Take into account the relevant cable length restrictions for your bus system,
which are dependent on the baud rate.
Содержание PW15AHi
Страница 2: ...English Page 3 0 24 Deutsch Seite 25 0 46 Fran ais Page 47 0 68...
Страница 69: ...69 PW15AHi A2526 4 0 en de fr HBM...
Страница 70: ...PW15AHi 70 A2526 4 0 en de fr HBM...