MGCplus−ML71BS6
A0851-3.1 en
HBM
Formats 2 and 3; 32-bit integers and 16-bit integers
You must also specify scaling here.
1
Use the shift key
to change to set−up mode and then press the
function key
. Use the cursor keys to select “Sources CAN1” and
confirm with
.
Sources CAN1
CHANNEL 3
CAN signal 1
System
Display
Parameters
Options
Gross
0
Source for
MGC channel
Signal
Scaling
Point 1
CAN signal 1
CAN signal 2
.
.
CAN signal n
Gross
Net
Store1
Store2
Combined PV
Point 2
dig
10
7680000
dig
1000
8
kN
kN
Channel 8 to
128
Fig. 4.3:
CAN1 / CAN 2 output signals dialogue window
2
Use the cursor keys
to select the “Source for MGC channel” selection
field. Use
to call the selection list and choose the desired signal.
Confirm with
.
3
Use the cursor keys
to select the MGC channel field and enter the
desired channel (e.g. 8.1).
4
Use the cursor keys
to select the “Signal” edit field. Use
to call the
selection list and choose the desired signal. Confirm with
.
5
Use the cursor keys
to select the “Point 1” edit field and enter the digit
value.
6
Use the cursor keys
to select the next edit field and enter the value of
the accompanying physical unit.
7
Use the cursor keys
to select the “Point 2” edit field and proceed in the
same way as for points 5 and 6.
8
Use the shift key
to change to measuring mode and confirm the
follow-up question with
.