MGCplus−ML71BS6

A0851-3.1 en

HBM

Formats 2 and 3; 32-bit integers and 16-bit integers

You must also specify scaling here.

1

Use the shift key 

 to change to set−up mode and then press the

function key 

. Use the cursor keys to select “Sources CAN1” and

confirm with 

.

Sources CAN1

CHANNEL 3

CAN signal 1

System

Display

Parameters

Options

Gross

0

Source for

MGC channel

Signal

Scaling
Point 1

CAN signal 1

CAN signal 2

.

.

CAN signal n

Gross

Net

Store1

Store2

Combined PV

Point 2

dig

10

7680000

dig

1000

8

kN

kN

Channel 8 to
128

Fig. 4.3:

CAN1 / CAN 2 output signals dialogue window

2

Use the cursor keys 

 to select the “Source for MGC channel” selection

field. Use 

 to call the selection list and choose the desired signal.

Confirm with 

.

3

Use the cursor keys 

 to select the MGC channel field and enter the

desired channel (e.g. 8.1).

4

Use the cursor keys 

 to select the “Signal” edit field. Use 

 to call the

selection list and choose the desired signal. Confirm with 

.

5

Use the cursor keys 

 to select the “Point 1” edit field and enter the digit

value.

6

Use the cursor keys 

 to select the next edit field and enter the value of

the accompanying physical unit.

7

Use the cursor keys 

 to select the “Point 2” edit field and proceed in the

same way as for points 5 and 6.

8

Use the shift key 

 to change to measuring mode and confirm the

follow-up question with 

.

Содержание ML71BS6

Страница 1: ...A0851 3 1 en MGCplus Measuring amplifier system CAN bus interface ML71BS6 Operating manual...

Страница 2: ......

Страница 3: ...Front panel 8 4 Parameterization 9 4 1 Settings CAN1 CAN2 9 4 1 1 Panel output signals CAN 1 11 5 Menu structure in set up mode 13 6 CAN protocol 14 6 1 Measurement signal transmission 14 6 1 1 Absolu...

Страница 4: ...untrained personnel Any person instructed to carry out installation commissioning maintenance or repair of the device must have read and understood the Operating Manual and in particular the technical...

Страница 5: ...71BS6 CAN bus module may be carried out without our express agreement Any modification shall exclude all liability on our part for any resulting damage In particular any repair or soldering work on mo...

Страница 6: ...s per CAN port scales them and outputs them at the CAN bus Two independent CAN interfaces per channel and two different physical layers ISO 11898 24V and a low speed link based on the Philips TJA1053...

Страница 7: ...systems female connectors CAN1 and CAN2 Both the CAN interfaces are used to output the MGCplus measurement data Each of the interfaces can be assigned up to 128 measured values Connectors male CAN1 CA...

Страница 8: ...LED caption Colour Meaning in Status mode CHAN yellow Channel selected ERROR WARN red error warning Rx1 not used Tx1 OVRN1 red yel low CAN protocol transmitted BUSERR1 red Bus error at CAN1 Rx2 yellow...

Страница 9: ...note that the maximum possible output rates can only be achieved if no signals have been selected at CAN port1 If a selected rate cannot be achieved the messages will be output as quickly as possible...

Страница 10: ...D Last ID Reset CAN1 port 640 641 Fig 4 1 CAN1 CAN2 dialogue window 2 Use the cursor keys to choose the baud rate selection field Use to call the selection list and choose the desired baud rate Confir...

Страница 11: ...signal 1 System Display Parameters Options Gross Source for MGC channel Signal CAN signal 1 CAN signal 2 CAN signal n Gross Net Store1 Store2 Combined PV 8 Fig 4 2 CAN1 CAN 2 output signals dialogue w...

Страница 12: ...e window 2 Use the cursor keys to select the Source for MGC channel selection field Use to call the selection list and choose the desired signal Confirm with 3 Use the cursor keys to select the MGC ch...

Страница 13: ...uring the initial start up or for a new measurement task User defiened dis play settings such as the required method of representing va lues function key as signment or the spe cification of channel n...

Страница 14: ...6 1 1 Absolute values in Float In each 8 byte CAN message 2 signals are transmitted as Float values 32 bit This is dealing with scaled measured values 6 1 2 In digits as Long In each CAN message 2 of...

Страница 15: ...red value 1 signed in tegers in digits from 30000 Measurement data messages CAN port 2 Absolute values in Float Identifier 8 byte data field 0x300 0x33F 4 byte measured value 1 absolute Float 4 byte m...

Страница 16: ...ue 2x 9 pin DSUB individual potential separation of supply and ground Measurement recording Number of signals at 1 Mbaud 11 bit Format INT16 Signals 1 8 Hz Output rate 2400 9 16 1200 17 24 800 25 32 6...

Страница 17: ...rd format mm Europe 100 x 160 Width mm 20 3 4 divs Receptacle DIN 41612 indirectly Analogue output The analogue output can display one of the max 128 input signals at a time Nominal rated voltage V 10...

Страница 18: ...18 MGCplus ML71BS6 A0851 3 1 en HBM...

Страница 19: ......

Страница 20: ...D 64293 Darmstadt Tel 49 6151 803 0 Fax 49 6151 8039100 Email support hbm com Internet www hbm com Modifications reserved All details describe our products in general form only They are not to be und...

Отзывы: